Combined analysis:

Analysis ID AN000694
Analysis type MS
Chromatography type GC
Chromatography system Agilent 6890
Column None
MS Type EI
MS instrument type Single quadrupole
MS instrument name Agilent 5973
Ion Mode POSITIVE
Units micromolar

Chromatography:

Chromatography ID:CH000502
Chromatography Summary:The derivatized extracts were analyzed with an Agilent 6890 gas chromatograph coupled with an Agilent 5973 mass selective detector. The 1-μL aliquots of the extracts were injected into a DB5-MS capillary column (30 m × 250 μm i.d., 0.25-μm film thickness; J&W Scientific, Folson, CA) using PTV injection (Gerstel CIS4 injector) in the splitless mode. The temperature of the PTV was 70 °C during injection, and 0.6 min after injection, the temperature was raised to 300 °C at a rate of 2 °C/s and held at 300 °C for 20 min. The initial GC oven temperature was 70 °C, 5 min after injection the GC oven temperature was increased with 5 °C/min to 320 °C and held for 5 min at 320 °C. Helium was used as a carrier gas and pressure programmed such that the helium flow was kept constant at a flow rate of 1.7 mL/min. Detection was achieved using MS detection in electron impact mode and full scan monitoring mode (m/z 15−800). The temperature of the ion source was set at 250 °C and that of the quadrupole at 200 °C.
Instrument Name:Agilent 6890
Column Name:None
Chromatography Type:GC