MB Sample ID: SA024600

Local Sample ID:ms5361-17
Subject ID:SU000504
Subject Type:Animal
Subject Species:Homo sapiens
Taxonomy ID:9606
Species Group:Human

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000749
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity
Column C18
MS Type ESI
MS instrument type Triple quadrupole
MS instrument name Thermo Quantum Ultra
Ion Mode POSITIVE
Units micromolar

Chromatography:

Chromatography ID:CH000537
Chromatography Summary:High resolution separation was done using an Acquity UPLC system, injecting 1 µl of derviatized solution, with a UPLC BEH C18 1.7 micron 2.1×150 mm column from Waters. Column flow was set to 400 µl/min with a gradient from 99.9%A to 98%B where buffer A is 1% acetonitrile in 0.1% formic acid and buffer B is 100% acetonitrile. A column temp of 43 degrees Celsius and a sample tray temp of 6% Celsius. Mass detection was completed on a TSQ Ultra Quantum from Thermo Finnigan running in ESI positive mode. A scan width of 0.002, scan time of 0.04 seconds per transition mass, collision energy of 25, collision gas pressure of 1.5 mTorr, tube lens value set to 90, monitoring a signature ion of the derivitized amines at m/z 171.04 by selected reaction monitoring.
Instrument Name:Waters Acquity
Column Name:C18
Column Temperature:43
Flow Gradient:99.9% A to 98% B
Flow Rate:400 µl/min
Solvent A:99% water/1% acetonitrile; 0.1% formic acid
Solvent B:100% acetonitrile
Chromatography Type:Reversed phase