Return to study ST000923 main page

MB Sample ID: SA054722

Local Sample ID:SM-7HNM8
Subject ID:SU000961
Subject Type:Human stool
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN001513 AN001514 AN001515 AN001516
Analysis type MS MS MS MS
Chromatography type HILIC HILIC Reversed phase Reversed phase
Chromatography system Shimadzu Nexera X2 Shimadzu Nexera X2 Shimadzu Nexera X2 Shimadzu Nexera X2
Column Atlantis HILIC (Waters; Milford,MA) Luna NH2 (Phenomenex; Torrance,CA) Waters ACQUITY UPLC BEH C18 Waters ACQUITY UPLC BEH C8 (1.7um)
MS Type ESI ESI ESI ESI
MS instrument type Orbitrap Orbitrap Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Plus Orbitrap Thermo Q Exactive Orbitrap Thermo Q Exactive Plus Orbitrap
Ion Mode POSITIVE NEGATIVE NEGATIVE POSITIVE
Units abundance abundance abundance abundance

Chromatography:

Chromatography ID:CH001066
Chromatography Summary:LC-MS samples were prepared from stool homogenates (10 μL) via protein precipitation with the addition of nine volumes of 74.9:24.9:0.2 v/v/v acetonitrile/methanol/formic acid containing stable isotope-labeled internal standards (valine-d8, Isotec; and phenylalanine-d8, Cambridge Isotope Laboratories; Andover, MA). The samples were centrifuged (10 min, 9,000 x g, 4°C), and the supernatants injected directly onto a 150 x 2 mm Atlantis HILIC column (Waters; Milford, MA). The column was eluted isocratically at a flow rate of 250 μL/min with 5% mobile phase A (10 mM ammonium formate and 0.1% formic acid in water) for 1 minute followed by a linear gradient to 40% mobile phase B (acetonitrile with 0.1% formic acid) over 10 minutes.
Instrument Name:Shimadzu Nexera X2
Column Name:Atlantis HILIC (Waters; Milford,MA)
Flow Gradient:The column was eluted isocratically with 5% A for 1 minute followed by a linear gradient to 40% B over 10 minutes.
Flow Rate:250 µL/min
Solvent A:100% water; 0.1% formic acid; 10 mM ammonium formate
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:HILIC
  
Chromatography ID:CH001067
Chromatography Summary:LC-MS samples were prepared from stool homogenates (30 μL) via protein precipitation with the addition of four volumes of 80% methanol containing inosine-15N4, thymine-d4 and glycocholate-d4 internal standards (Cambridge Isotope Laboratories; Andover, MA). The samples were centrifuged (10 min, 9,000 x g, 4°C) and the supernatants were injected directly onto a 150 x 2.0 mm Luna NH2 column (Phenomenex; Torrance, CA). The column was eluted at a flow rate of 400 μL/min with initial conditions of 10% mobile phase A (20 mM ammonium acetate and 20 mM ammonium hydroxide in water) and 90% mobile phase B (10 mM ammonium hydroxide in 75:25 v/v acetonitrile/methanol) followed by a 10 min linear gradient to 100% mobile phase A.
Instrument Name:Shimadzu Nexera X2
Column Name:Luna NH2 (Phenomenex; Torrance,CA)
Flow Gradient:The column was eluted with initial conditions of 10% mobile phase A and 90% mobile phase B followed by a 10 min linear gradient to 100% mobile phase A.
Flow Rate:400 µL/min
Solvent A:100% water; 20 mM ammonium acetate; mM ammonium hydroxide
Solvent B:75% acetonitrile/25% methanol; 10 mM ammonium hydroxide
Chromatography Type:HILIC
  
Chromatography ID:CH001068
Chromatography Summary:Stool homogenates (30 μL) were extracted using 90 μL of methanol containing PGE2-d4 as an internal standard (Cayman Chemical Co.; Ann Arbor, MI) and centrifuged (10 min, 9,000 x g, 4°C). The supernatants (10 μL) were injected onto a 150 x 2.1 mm ACQUITY BEH C18 column (Waters; Milford, MA). The column was eluted isocratically at a flow rate of 450 μL/min with 20% mobile phase A (0.01% formic acid in water) for 3 minutes followed by a linear gradient to 100% mobile phase B (0.01% acetic acid in acetonitrile) over 12 minutes.
Instrument Name:Shimadzu Nexera X2
Column Name:Waters ACQUITY UPLC BEH C18
Flow Gradient:The column was eluted isocratically at a flow rate of 450 µL/min with 20% A for 3 minutes followed by a linear gradient to 100% B over 12 minutes.
Flow Rate:450 µL/min
Solvent A:100% water; 0.01% formic acid
Solvent B:100% acetonitrile; 0.01% acetic acid
Chromatography Type:Reversed phase
  
Chromatography ID:CH001069
Chromatography Summary:Lipids (polar and nonpolar) were extracted from stool homogenates (10 μL) using 190 μL of isopropanol containing 1-dodecanoyl-2-tridecanoyl-sn-glycero-3-phosphocholine as an internal standard (Avanti Polar Lipids; Alabaster, AL). After centrifugation (10 min, 9,000 x g, ambient temperature), supernatants (10 μL) were injected directly onto a 100 x 2.1 mm ACQUITY BEH C8 column (1.7 μm; Waters; Milford, MA). The column was eluted at a flow rate of 450 μL/min isocratically for 1 minute at 80% mobile phase A (95:5:0.1 vol/vol/vol 10 mM ammonium acetate/methanol/acetic acid), followed by a linear gradient to 80% mobile-phase B (99.9:0.1 vol/vol methanol/acetic acid) over 2 minutes, a linear gradient to 100% mobile phase B over 7 minutes, and then 3 minutes at 100% mobile-phase B.
Instrument Name:Shimadzu Nexera X2
Column Name:Waters ACQUITY UPLC BEH C8 (1.7um)
Flow Gradient:The column was eluted isocratically for 1 minute at 80% mobile phase A, followed by a linear gradient to 80% mobile-phase B over 2 minutes, a linear gradient to 100% mobile phase B over 7 minutes, and then 3 minutes at 100% mobile-phase B.
Flow Rate:450 µL/min
Solvent A:95% water/5% methanol; 0.1% acetic acid; 10 mM ammonium acetate
Solvent B:100% methanol; 0.1% acetic acid
Chromatography Type:Reversed phase
  logo