Return to study ST001240 main page

MB Sample ID: SA089954

Local Sample ID:HTP_4neg_D21_82
Subject ID:SU001308
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Age Or Age Range:0.5 - 76.5
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002059 AN002060
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Phenomenex Kinetex C18 (150 x 2.1mm,2.6um) Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode NEGATIVE POSITIVE
Units relative abundance relative abundance

Chromatography:

Chromatography ID:CH001497
Chromatography Summary:UHPLC-MS metabolomics analyses were performed using a Vanquish UHPLC system coupled online to a Q Exactive mass spectrometer (Thermo Fisher, Bremen, Germany). Samples were resolved over a Kinetex C18 column (2.1x150 mm, 1.7μm; Phenomenex, Torrance, CA, USA) at 25°C using a 3-minute isocratic flow rate at 250μL/minute at 0% B (A: 95% water/5% acetonitrile, 5mM NH4OAc) for positive mode. To monitor possible technical variability, aliquots of each of the individual samples were combined to make technical replicates, which were run after every 15 samples. Additionally, in each experiment, several lysis solution aliquots were run as blanks for artifact identification. The Q Exactive mass spectrometer (Thermo Fisher Scientific, San Jose, CA, USA) was operated in negative ion mode using electrospray ionization, scanning in Full MS mode (2μscans) from 65 to 900 m/z at 70,000 resolution, with 4 kV spray voltage, 15 sheath gas, 5 auxiliary gas. MS analysis and data elaboration were performed as described74,75. Calibration was performed prior to analysis using the PierceTM Negative Ion Calibration Solution (Thermo Fisher Scientific).
Methods Filename:4MMneg_method.pdf
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Column Temperature:45
Flow Gradient:0-100% B
Flow Rate:450uL/min
Injection Temperature:4
Solvent A:95% water/5% acetonitrile; 5 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 5 mM ammonium acetate
Analytical Time:4min
Capillary Voltage:4kV
Washing Buffer:10% methanol
Chromatography Type:Reversed phase
  
Chromatography ID:CH001498
Chromatography Summary:UHPLC-MS metabolomics analyses were performed using a Vanquish UHPLC system coupled online to a Q Exactive mass spectrometer (Thermo Fisher, Bremen, Germany). Samples were resolved over a Kinetex C18 column (2.1x150 mm, 1.7μm; Phenomenex, Torrance, CA, USA) at 45°C using a 4 minute gradient at 450μL/minute from 0-100% B (A: 95% water/5% acetonitrile, 5mM NH4OAc; B: 95% acetonitrile/5% water, 5mM NH4OAc) for negative mode. To monitor possible technical variability, aliquots of each of the individual samples were combined to make technical replicates, which were run after every 15 samples. Additionally, in each experiment, several lysis solution aliquots were run as blanks for artifact identification.
Methods Filename:4MMpos_method.pdf
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Column Temperature:45
Flow Gradient:5-95% B
Flow Rate:450uL/min
Injection Temperature:4
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Analytical Time:4min
Capillary Voltage:4kV
Washing Buffer:10% methanol
Chromatography Type:Reversed phase
  logo