Return to study ST001265 main page
MB Sample ID: SA092007
Local Sample ID: | MeOH-H2O |
Subject ID: | SU001333 |
Subject Type: | Cultured cells |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Gender: | Female |
Cell Biosource Or Supplier: | ATCC |
Cell Strain Details: | MCF-7 |
Cell Passage Number: | 30 |
Species Group: | Homo Sapien |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN002102 |
---|---|
Analysis type | MS |
Chromatography type | GC |
Chromatography system | Shimadzu GCMS-QP2010 Ultra |
Column | Restek Rtx-5Sil (30m x 0.25mm,0.25um) |
MS Type | EI |
MS instrument type | Single quadrupole |
MS instrument name | Shimadzu QP2010 Ultra |
Ion Mode | POSITIVE |
Units | Peak Area |
Chromatography:
Chromatography ID: | CH001534 |
Chromatography Summary: | A GC/MS-QP 2010 Ultra System (Shimadzu, Kyoto, Japan) was employed for the metabolomic analysis, along with LabSolutions GC-MS software (ver). A Restek Rtx®-5ms column (30.0 m × 0.25 mm, 0.25 µm) was utilized for separation of the metabolites. Helium (99.9%) was used as the carrier gas at the flow rate of 1.0 mL/min. The initial oven temperature was set at 60 °C and was held at this temperature for 2 minutes, then raised to 310 °C by 50 °C/min and held at this temperature during the analysis. Both the auxiliary temperature at the interface and the ionization temperature were 250 °C. Metabolites were analysed in full scan mode within the range of 50 – 650 amu. Total volume of 10 µL was injected in splitless mode, by employing AOC-20i Auto Injector (Shimadzu, Kyoto, Japan). GC total ion chromatograms (TIC) and fragmentation patterns of the metabolites identified using NIST/EPA/NIH Mass Spectral Library (NIST 14). Run time for each sample was 43.67 min. |
Methods Filename: | PAH scan |
Instrument Name: | Shimadzu GCMS-QP2010 Ultra |
Column Name: | Restek Rtx-5Sil (30m x 0.25mm,0.25um) |
Column Pressure: | 57.4 |
Column Temperature: | 60 |
Flow Rate: | 1 |
Injection Temperature: | 250 |
Retention Time: | many retention time |
Sample Injection: | 1 |
Solvent A: | Pyridine |
Analytical Time: | 43.67 |
Oven Temperature: | 60 |
Time Program: | 43.67 |
Transferline Temperature: | 250 |
Strong Wash Solvent Name: | Acetone |
Strong Wash Volume: | 10 uL |
Sample Syringe Size: | 10 uL |
Chromatography Type: | GC |