MB Sample ID: SA098630

Local Sample ID:S28R3
Subject ID:SU001430
Subject Type:Other
Subject Species:Ilyonectria mors-panacis;Ilyonectria rufa;Ilyonectria robusta;Ilyonectria torresensis;Ilyonectria estremocensis;Neonectria obtusispora
Taxonomy ID:1079131;1079113;1079257;1199091;1079267;78417

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002255 AN002256 AN002257 AN002258 AN002259
Analysis type MS MS MS MS MS
Chromatography type Reversed phase Reversed phase HILIC HILIC HILIC
Chromatography system Waters Acquity I-Class Waters Acquity I-Class Waters Acquity I-Class Waters Acquity I-Class Waters Acquity I-Class
Column Waters Acquity UPLC HSS Cyano (CN) ( 50 x 2.1mm,1.8um) Waters Acquity UPLC HSS Cyano (CN) ( 50 x 2.1mm,1.8um) SeQuant ZIC-pHILIC (150 x 2.1mm,5um) SeQuant ZIC-pHILIC (150 x 2.1mm,5um) SeQuant ZIC-pHILIC (150 x 2.1mm,5um)
MS Type ESI ESI ESI ESI ESI
MS instrument type Triple quadrupole Orbitrap Triple quadrupole Triple quadrupole Orbitrap
MS instrument name Waters Xevo-TQ-S Thermo Q Exactive HF hybrid Orbitrap Waters Xevo-TQ-S Waters Xevo-TQ-S Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE POSITIVE POSITIVE NEGATIVE POSITIVE
Units Normalized Peak Area Per L Seawater Filtered Normalized Peak Area Per L Seawater Filtered Normalized Peak Area Per L Seawater Filtered Normalized Peak Area Per L Seawater Filtered Normalized Peak Area per L of SW filtered

Chromatography:

Chromatography ID:CH001657
Chromatography Summary:RP
Methods Filename:Ingalls_Metabolomics_MS_Parameters_2015.pdf
Instrument Name:Waters Acquity I-Class
Column Name:Waters Acquity UPLC HSS Cyano (CN) ( 50 x 2.1mm,1.8um)
Column Temperature:35 C
Flow Gradient:5% B for 2 minutes, ramped to 100% B over 16 minutes, held at 100% B for 2 minutes, and equilibrated at 5% B for 5 minutes
Flow Rate:0.4 mL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Analytical Time:20 min
Preconditioning:equilibrated at the starting conditions for at least 10 minutes; several water blanks were run before
Chromatography Type:Reversed phase
  
Chromatography ID:CH001658
Chromatography Summary:HILIC
Methods Filename:Ingalls_Metabolomics_MS_Parameters_2015.pdf
Chromatography Comments:In marine samples, a major salt peak elutes at approximately 23 minutes, To improve the performance of the HILIC column, we maintained the same injection volume, kept the instrument running water blanks between samples as necessary.
Instrument Name:Waters Acquity I-Class
Column Name:SeQuant ZIC-pHILIC (150 x 2.1mm,5um)
Column Temperature:30 C
Flow Gradient:100% A for 2 minutes, ramped to 100% B over 18 minutes, held at 100% B for 5 minutes, and equilibrated at 100% A for 25 minutes
Flow Rate:0.15 mL/min
Solvent A:85% acetonitrile/15% water; 10 mM ammonium carbonate
Solvent B:40% acetonitrile/60% water; 10 mM ammonium carbonate
Analytical Time:25 min
Preconditioning:, the column was equilibrated at the starting conditions for at least 30 minutes
Chromatography Type:HILIC