Return to study ST001641 main page

MB Sample ID: SA151254

Local Sample ID:12 T_065
Subject ID:SU001718
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002686
Analysis type MS
Chromatography type Reversed phase
Chromatography system Agilent
Column Waters ACQUITY UPLC CSH C18
MS Type EI
MS instrument type QTOF
MS instrument name Agilent 6550 QTOF
Ion Mode POSITIVE
Units counts per second

Chromatography:

Chromatography ID:CH001980
Chromatography Summary:The LC/QTOFMS analyses are performed using an Agilent 1290 Infinity LC system (G4220A binary pump, G4226A autosampler, and G1316C Column Thermostat) coupled to either an Agilent 6530 (positive ion mode) or an Agilent 6550 mass spectrometer equipped with an ion funnel (iFunnel) (negative ion mode). Lipids are separated on an Acquity UPLC CSH C18 column (100 x 2.1 mm; 1.7 µm) maintained at 65°C at a flow-rate of 0.6 mL/min. Solvent pre-heating (Agilent G1316) was used. The mobile phases consist of 60:40 acetonitrile:water with 10 mM ammonium formate and 0.1% formic acid (A) and 90:10 propan-2-ol:acetonitrile with 10 mM ammonium formate and 0.1% formic acid. The gradient is as follows: 0 min 85% (A); 0–2 min 70% (A); 2–2.5 min 52% (A); 2.5–11 min 18% (A); 11–11.5 min 1% (A); 11.5–12 min 1% (A); 12–12.1 min 85% (A); 12.1–15 min 85% (A). A sample volume of 3 µL is used for the injection. Sample temperature is maintained at 4°C in the autosampler.
Instrument Name:Agilent
Column Name:Waters ACQUITY UPLC CSH C18
Column Temperature:65
Flow Gradient:0 min 85% (A); 0-2 min 70% (A); 2-2.5 min 52% (A); 2.5-11 min 18% (A); 11-11.5 min 1% (A); 11.5-12 min 1% (A); 12-12.1 min 85% (A); 12.1-15 min 85% (A).
Flow Rate:0.6 mL/min
Solvent A:60% acetonitrile/40% water; 0.1% formic acid; 10 mM ammonium formate
Solvent B:90% isopropanol/10% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Chromatography Type:Reversed phase