Return to study ST001671 main page

MB Sample ID: SA153570

Local Sample ID:ncbimw_0006
Subject ID:SU001748
Subject Type:Bacteria
Subject Species:Bacteria
Subject Comments:178 bacterial strains

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002727 AN002728 AN002729
Analysis type MS MS MS
Chromatography type Reversed phase Reversed phase HILIC
Chromatography system Agilent 1290 Infinity II Agilent 1290 Infinity II Agilent 1290 Infinity II
Column Waters Acquity BEH C18 (100 x 2mm,1.7um) Waters Acquity BEH C18 (100 x 2mm,1.7um) Waters Acquity BEH Amide (150 x 2.1mm,1.7um)
MS Type ESI ESI ESI
MS instrument type QTOF QTOF QTOF
MS instrument name Agilent 6540 QTOF Agilent 6540 QTOF Agilent 6540 QTOF
Ion Mode POSITIVE NEGATIVE POSITIVE
Units counts (area) counts (area) counts (area)

Chromatography:

Chromatography ID:CH002013
Chromatography Summary:Please see `methods.pdf` associated with this submission for full details. In short, this analysis is processed data associated with standard reverse phase LCMS with positive ionization.
Instrument Name:Agilent 1290 Infinity II
Column Name:Waters Acquity BEH C18 (100 x 2mm,1.7um)
Column Temperature:40
Flow Rate:350 uL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% methanol; 0.1% formic acid
Capillary Voltage:4000
Chromatography Type:Reversed phase
  
Chromatography ID:CH002014
Chromatography Summary:Please see `methods.pdf` associated with this submission for full details. In short, this analysis is processed data associated with standard reverse phase LCMS with negative ionization.
Instrument Name:Agilent 1290 Infinity II
Column Name:Waters Acquity BEH C18 (100 x 2mm,1.7um)
Column Temperature:40
Flow Rate:350 uL/min
Solvent A:100% water; 6.5 mM ammonium bicarbonate
Solvent B:95% methanol/5% water; 6.5 mM ammonium bicarbonate
Capillary Voltage:4000
Chromatography Type:Reversed phase
  
Chromatography ID:CH002015
Chromatography Summary:Please see `methods.pdf` associated with this submission for full details. In short, this analysis is processed data associated with HILIC LCMS with positive ionization.
Instrument Name:Agilent 1290 Infinity II
Column Name:Waters Acquity BEH Amide (150 x 2.1mm,1.7um)
Column Temperature:40
Flow Rate:400 uL/min
Solvent A:100% water; 0.125% formic acid; 10 mM ammonium formate
Solvent B:95% acetonitrile/5% water; 0.125% formic acid; 10 mM ammonium formate
Capillary Voltage:4000
Chromatography Type:HILIC
  logo