Return to study ST001720 main page

MB Sample ID: SA162020

Local Sample ID:QC1-3
Subject ID:SU001797
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002803
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Vanquish
Column SeQuant ZIC-pHILIC (150 x 2.1mm,5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF-X Orbitrap
Ion Mode UNSPECIFIED
Units Peak Area

Chromatography:

Chromatography ID:CH002071
Chromatography Summary:Hydrophilic interaction liquid chromatography (HILIC) was performed at 0.2 ml/min on a ZIC-pHILIC column (2.1 mm × 150 mm, EMD Millipore) at 45 °C. Solvent A was 20 mM ammonium carbonate, 0.1% ammonium hydroxide, pH 9.2, and solvent B was acetonitrile. The gradient was 85% B for 2 min, 85% B to 20% B over 15 min, 20% B to 85% B over 0.1 min, and 85% B for 8.9 min. The autosampler was held at 4 °C. For each analysis, 4 µl of sample was injected.
Instrument Name:Thermo Vanquish
Column Name:SeQuant ZIC-pHILIC (150 x 2.1mm,5um)
Flow Gradient: 85% B for 2 min, 85% B to 20% B over 15 min, 20% B to 85% B over 0.1 min, and 85% B for 8.9 min.
Solvent A:100% water; 20 mM ammonium carbonate; 0.1% ammonium hydroxide, pH 9.3
Solvent B:100% acetonitrile
Chromatography Type:HILIC