Return to study ST001732 main page

MB Sample ID: SA162639

Local Sample ID:crtl3
Subject ID:SU001809
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002819 AN002820
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Phenomenex Kinetex C18 (150 x 2.1mm,2.6um) Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units Relative Abundance Relative Abundance

Chromatography:

Chromatography ID:CH002084
Chromatography Summary:Positive Mode 10 μL of water and methanol soluble fractions were run through a Kinetex C18 1.7 μm, 100 x 2.1 mm (Phenomenex) reversed phase column (Positive ion mode - phase A: water, 0.1 % formic acid; B: acetonitrile, 0.1 % formic acid) via an ultra-high performance chromatographic system (UHPLC - Vanquish, Thermo Fisher). UHPLC was coupled in line with a high-resolution quadrupole Orbitrap instrument run in positive polarity mode (QExactive, Thermo Fisher) at 70,000 resolution (at 200 m/z). Gradients and other technical parameters and the variants employed herein have been extensively described.
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Solvent A:100% water; 0.1 % formic acid
Solvent B:100% acetonitrile, 0.1 % formic acid
Chromatography Type:Reversed phase
  
Chromatography ID:CH002085
Chromatography Summary:Negative Mode 10 μL of water and methanol soluble fractions were run through a Kinetex C18 1.7 μm, 100 x 2.1 mm (Phenomenex) reversed phase column (Negative ion mode – phase A: 1 mM NH4Ac 95:5 water : acetonitrile; phase B: 1 mM NH4Ac 95:5 acetonitrile : water) via an ultra-high performance chromatographic system (UHPLC - Vanquish, Thermo Fisher). UHPLC was coupled in line with a high-resolution quadrupole Orbitrap instrument run in negative polarity mode (QExactive, Thermo Fisher) at 70,000 resolution (at 200 m/z). Gradients and other technical parameters and the variants employed herein have been extensively described.
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Solvent A:95% water/5% acetonitrile; 1 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 1 mM ammonium acetate
Chromatography Type:Reversed phase
  logo