Return to study ST001774 main page

MB Sample ID: SA164540

Local Sample ID:2020_04_15_AM_HLICNeg15_HBE85_4_C
Subject ID:SU001851
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Not applicable

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002881
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Vanquish
Column ThermoFisher Syncronis HILIC UHPLC (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Single quadrupole
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode NEGATIVE
Units pmoles/l

Chromatography:

Chromatography ID:CH002136
Chromatography Summary:General metabolomic run were performed on a Q Exactive™ HF Hybrid Quadrupole-Orbitrap™ Mass Spectrometer (ThermoFisher) coupled to a Vanquish™ UHPLC System (ThermoFisher). Chromatographic separation was achieved on a Syncronis HILIC UHPLC column (2.1 mm x 100 mm x 1.7 mm, ThermoFisher) using a binary solvent system at a flow rate of 600 mL/min. Solvent A, 20 mM ammonium formate pH 3.0 in mass spectrometry grade H2O; Solvent B, mass spectrometry grade acetonitrile with 0.1% formic acid (% v/v). A sample injection volume of 2 mL was used.
Instrument Name:Thermo Vanquish
Column Name:ThermoFisher Syncronis HILIC UHPLC (100 x 2.1mm,1.7um)
Solvent A:100% water; 20 mM ammonium formate, pH 3
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:HILIC
  logo