Combined analysis:

Analysis ID AN002856
Analysis type MS
Chromatography type HILIC
Chromatography system Agilent 1290 Infinity II
Column Waters Acquity BEH Amide (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type QTOF
MS instrument name Agilent 6545
Ion Mode POSITIVE
Units Total Intensity Units

Chromatography:

Chromatography ID:CH002383
Chromatography Summary:HILIC separations were performed on a Waters Acquity UPLC BEH amide column (2.1 x 100 mm 1.7 m) with a matching Vanguard precolumn. The flow rate was set to 0.3 mL/min and mobile phases consisted of (A) 95:5 acetonitrile/water with 0.125% v/v formic acid and 10 mM ammonium formate and (B) 95:5 water/acetonitrile with 0.125% v/v formic acid and 10 mM ammonium formate. Gradient specified in protocol file.
Methods Filename:2 mm amide pos MS2 70%H2O 120min_PC6.m
Instrument Name:Agilent 1290 Infinity II
Column Name:Waters Acquity BEH Amide (100 x 2.1mm,1.7um)
Column Pressure:300 bar
Column Temperature:55
Flow Rate:0.3 mL/min
Sampling Cone:175 V
Solvent A:95% acetonitrile/5% water; 0.125% formic acid; 10 mM ammonium formate
Solvent B:5% acetonitrile/95% water; 0.125% formic acid; 10 mM ammonium formate
Analytical Time:120 min
Capillary Voltage:3500 V
Weak Wash Solvent Name:85:15 acetonitrile:water
Target Sample Temperature:4
Sample Loop Size:20 uL
Chromatography Type:HILIC