Return to study ST002016 main page

MB Sample ID: SA188686

Local Sample ID:Pat_21_NEG
Subject ID:SU002097
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003284 AN003285
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Thermo Accucore C18 (100 x 2.1mm,2.6um) Thermo Accucore C18 (100 x 2.1mm,2.6um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Orbitrap ID-X tribrid Thermo Orbitrap ID-X tribrid
Ion Mode POSITIVE NEGATIVE
Units intensity intensity

Chromatography:

Chromatography ID:CH002425
Chromatography Summary:Samples were injected in randomized fashion via a Thermo Vanquish UHPLC and separation of the polar metabolites was achieved using Thermo Accucore C18 column (Thermo Scientific; 2.1 x 100 mm; 1.5 µm) maintained at 30 °C. The injection volume was 10 µL. For the 15-minute gradient, the standard mobile phase for RPLC was A = 0.1% formic acid in water and B = 0.1% formic acid in methanol. The linear elution gradient was as follows: 0-8.0 minutes at 50% B, 8.0 – 13.0 minutes held at 98% B, 13.1 to 15.0 minutes revert to 0% B to re-equilibration for next injection at a flow rate of 0.25 mL/min.
Instrument Name:Thermo Vanquish
Column Name:Thermo Accucore C18 (100 x 2.1mm,2.6um)
Column Temperature:30C
Flow Gradient:0-8 minutes 50% B, 8 - 13 minutes held at 98% B, 13.1 to 15.0 minutes revert to 0% B to re-equilibration for next injection
Flow Rate:0.25 mL/min
Solvent A:100% water; 0.1% formic acid
Solvent B:100% methanol; 0.1% formic acid
Chromatography Type:Reversed phase
  logo