MB Sample ID: SA237321

Local Sample ID:LA-2
Subject ID:SU002460
Subject Type:Cultured cells
Subject Species:Mus musculus
Taxonomy ID:10090

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003866
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Dionex Ultimate 3000
Column Waters XBridge Amide (100 x 4.6mm,3.5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Plus Orbitrap
Ion Mode UNSPECIFIED
Units peak area

Chromatography:

Chromatography ID:CH002864
Chromatography Summary:The HPLC (Ultimate 3000 UHPLC) is coupled to QE-MS (Thermo Scientific) for metabolite separation and detection. An Xbridge amide column (100 × 2.1 mm i.d., 3.5 μm; Waters) is employed for compound separation at room temperature. The mobile phase A is water containing 5 mM ammonium acetate (pH 6.8), and the mobile phase B is acetonitrile. The linear gradient used is as follows: 0 min, 85% B; 1.5 min, 85% B, 5.5 min, 35% B; 10 min, 35% B, 10.5 min, 35% B, 14.5 min, 35% B, 15 min, 85% B, and 20 min, 85% B. The flow rate was 0.15 mL/min from 0 to 10 min and 15 to 20 min and 0.3 mL/min from 10.5 to 14.5 min. The QE-MS is equipped with a HESI probe with related parameters set as below: heater temperature, 120 °C; sheath gas, 30; auxiliary gas, 10; sweep gas, 3; spray voltage, 3.0 kV for the positive mode and 2.5 kV for the negative mode; capillary temperature, 320 °C; S-lens, 55; scan range (m/z): 70 to 900 for pos mode (1.31 to 12.5 min) and neg mode (1.31 to 6.6 min) and 100 to 1000 for neg mode (6.61 to 12.5 min); resolution: 70000; automated gain control (AGC), 3 × 106 ions. Customized mass calibration was performed before data acquisition.
Methods Filename:The_HPLC_methods.docx
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:Waters XBridge Amide (100 x 4.6mm,3.5um)
Flow Gradient:The linear gradient used is as follows: 0 min, 85% B; 1.5 min, 85% B, 5.5 min, 35% B; 10 min, 35% B, 10.5 min, 35% B, 14.5 min, 35% B, 15 min, 85% B, and 20 min, 85% B.
Flow Rate:0.15 mL/min from 0 to 10 min and 15 to 20 min and 0.3 mL/min from 10.5 to 14.5 min.
Solvent A:100% water; 5 mM ammonium acetate, pH 6.8
Solvent B:100% acetonitrile
Chromatography Type:HILIC