Return to study ST002477 main page

MB Sample ID: SA248008

Local Sample ID:PMN_32
Subject ID:SU002567
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004046
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Scientific QExactive HF-X hybrid quadrupole orbitrap high-resolution mass spectrometer (HRMS)
Column Millipore ZIC-pHILIC column (5 mm, 2.1×150 mm)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF-X Orbitrap
Ion Mode UNSPECIFIED
Units MS reading

Chromatography:

Chromatography ID:CH002993
Chromatography Summary:Liquid chromatography was performed using a Millipore ZIC-pHILIC column (5 mm, 2.1×150 mm) with a binary solvent system of 10 mM ammonium acetate in water, pH 9.8 (solvent A) and acetonitrile (solvent B) with a constant flow rate of 0.25 mL/min. The column was equilibrated with 90% solvent B. The liquid chromatography gradient was: 0–15 minutes linear ramp from 90% B to 30% B; 15–18 minutes isocratic flow of 30% B; 18–19 minutes linear ramp from 30% B to 90% B; 19–27 column regeneration with isocratic flow of 90% B.
Instrument Name:Thermo Scientific QExactive HF-X hybrid quadrupole orbitrap high-resolution mass spectrometer (HRMS)
Column Name:Millipore ZIC-pHILIC column (5 mm, 2.1×150 mm)
Column Temperature:40
Flow Gradient:0–15 minutes linear ramp from 90% B to 30% B; 15–18 minutes isocratic flow of 30% B; 18–19 minutes linear ramp from 30% B to 90% B; 19–27 column regeneration with isocratic flow of 90% B.
Flow Rate:0.25 mL/min
Solvent A:100% water; 10 mM ammonium acetate, pH 9.8
Solvent B:100% acetonitrile
Chromatography Type:HILIC
  logo