Return to study ST002477 main page
MB Sample ID: SA248041
Local Sample ID: | PMN_50 |
Subject ID: | SU002567 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004046 |
---|---|
Analysis type | MS |
Chromatography type | HILIC |
Chromatography system | Thermo Scientific QExactive HF-X hybrid quadrupole orbitrap high-resolution mass spectrometer (HRMS) |
Column | Millipore ZIC-pHILIC column (5 mm, 2.1×150 mm) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive HF-X Orbitrap |
Ion Mode | UNSPECIFIED |
Units | MS reading |
Chromatography:
Chromatography ID: | CH002993 |
Chromatography Summary: | Liquid chromatography was performed using a Millipore ZIC-pHILIC column (5 mm, 2.1×150 mm) with a binary solvent system of 10 mM ammonium acetate in water, pH 9.8 (solvent A) and acetonitrile (solvent B) with a constant flow rate of 0.25 mL/min. The column was equilibrated with 90% solvent B. The liquid chromatography gradient was: 0–15 minutes linear ramp from 90% B to 30% B; 15–18 minutes isocratic flow of 30% B; 18–19 minutes linear ramp from 30% B to 90% B; 19–27 column regeneration with isocratic flow of 90% B. |
Instrument Name: | Thermo Scientific QExactive HF-X hybrid quadrupole orbitrap high-resolution mass spectrometer (HRMS) |
Column Name: | Millipore ZIC-pHILIC column (5 mm, 2.1×150 mm) |
Column Temperature: | 40 |
Flow Gradient: | 0–15 minutes linear ramp from 90% B to 30% B; 15–18 minutes isocratic flow of 30% B; 18–19 minutes linear ramp from 30% B to 90% B; 19–27 column regeneration with isocratic flow of 90% B. |
Flow Rate: | 0.25 mL/min |
Solvent A: | 100% water; 10 mM ammonium acetate, pH 9.8 |
Solvent B: | 100% acetonitrile |
Chromatography Type: | HILIC |