Return to study ST001700 main page

MB Sample ID: SA158230

Local Sample ID:6_serum_006
Subject ID:SU001777
Subject Type:Mammal
Subject Species:Mus musculus
Taxonomy ID:10090

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002771
Analysis type MS
Chromatography type GC
Chromatography system Leco Pegasus IV GC
Column Restek Rtx-5Sil (30m x 0.25mm,0.25um)
MS Type EI
MS instrument type GC-TOF
MS instrument name Leco Pegasus IV TOF
Ion Mode UNSPECIFIED
Units normalized peak height

MS:

MS ID:MS002568
Analysis ID:AN002771
Instrument Name:Leco Pegasus IV TOF
Instrument Type:GC-TOF
MS Type:EI
MS Comments:A Leco Pegasus IV time of flight mass spectrometer is controlled by the Leco ChromaTOF software vs. 2.32 (St. Joseph, MI). The transfer line temperature between gas chromatograph and mass spectrometer is set to 280°C. Electron impact ionization at 70V is employed with an ion source temperature of 250°C. Acquisition rate is 17 spectra/second, with a scan mass range of 85-500 Da.
Ion Mode:UNSPECIFIED
  logo