Analysis ID AN003295
Analysis type MS
Chromatography type GC
Chromatography system Leco Pegasus IV GC
Column Restek Rtx-5Sil (30m x 0.25mm,0.25um)
MS Type EI
MS instrument type GC-TOF
MS instrument name Leco Pegasus IV TOF
Ion Mode UNSPECIFIED
Units normalized peak height

MS:

MS ID:MS003065
Analysis ID:AN003295
Instrument Name:Leco Pegasus IV TOF
Instrument Type:GC-TOF
MS Type:EI
MS Comments:Gas Chromatography conditions: A 30 m long, 0.25 mm i.d. Rtx-5Sil MS column (0.25 μm 95% dimethyl 5% diphenyl polysiloxane film) with additional 10 m integrated guard column is used (Restek, Bellefonte PA). 99.9999% pure Helium with built-in purifier (Airgas, Radnor PA) is set at constant flow of 1 ml/min. The oven temperature is held constant at 50°C for 1 min and then ramped at 20°C/min to 330°C at which it is held constant for 5 min. Mass spectrometer settings: A Leco Pegasus IV time of flight mass spectrometer is controlled by the Leco ChromaTOF software vs. 2.32 (St. Joseph, MI). The transfer line temperature between gas chromatograph and mass spectrometer is set to 280°C. Electron impact ionization at 70V is employed with an ion source temperature of 250°C. Acquisition rate is 17 spectra/second, with a scan mass range of 85-500 Da.