Return to study ST002028 main page

MB Sample ID: SA190724

Local Sample ID:RS-01685857
Subject ID:SU002110
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003298
Analysis type MS
Chromatography type HILIC
Chromatography system Vanquish
Column Waters Acquity UPLC BEH amide,(150 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF hybrid Orbitrap
Ion Mode POSITIVE
Units normalized peak height

MS:

MS ID:MS003068
Analysis ID:AN003298
Instrument Name:Thermo Q Exactive HF hybrid Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:A Thermo Q-Exactive HF Orbitrap MS instrument was operated in positive and negative ESI mdoes respectively with the following parameters: mass range 60−900 m/z; spray voltage 3.6kV (ESI+) and −3kV (ESI−), sheath gas (nitrogen) flow rate 60 units; auxiliary gas (nitrogen) flow rate 25 units, capillary temperature 320 ◦C, full scan MS1 mass resolving power 60,000, data-dependent MSMS (dd-MSMS) 4 scans per cycle, normalized collision energy at 20%, 30%, and 40%, dd-MSMS mass resolving power 15,000. Thermo Xcalibur 4.0.27.19 was used for data acquisition and analysis. Instruments was tuned and calibrated by manufacturer’s recommendations.
Ion Mode:POSITIVE