Return to study ST002058 main page

MB Sample ID: SA193928

Local Sample ID:CTL-7_R3+
Subject ID:SU002140
Subject Type:Mammal
Subject Species:Mus musculus
Taxonomy ID:10090

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003350 AN003351
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system QExactive QExactive
Column Kinetix C18 (100 x 2.1mm) Kinetix C18 (100 x 2.1mm)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units AU AU

MS:

MS ID:MS003119
Analysis ID:AN003350
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The Q Exactive mass spectrometer (ThermoFisher) was operated independently in positive or negative ion mode, scanning in Full MS mode (2 scans) from 60 to 900 m/z at 70,000 resolution, with 4 kV spray voltage, 45 sheath gas, 15 auxiliary gas. Calibration was performed prior to analysis using the PierceTM Positive and Negative Ion Calibration Solutions (ThermoFisher). Acquired data was then converted from raw to mzXML file format using Mass Matrix (Cleveland, OH). Samples were analyzed in randomized order with a technical mixture injected periodically through analysis to qualify instrument performance. Metabolite assignments, isotopologue distributions, and correction for expected natural abundances of deuterium, 13C, and 15N isotopes were performed using MAVEN (Princeton, NJ)
Ion Mode:POSITIVE
  
MS ID:MS003120
Analysis ID:AN003351
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The Q Exactive mass spectrometer (ThermoFisher) was operated independently in positive or negative ion mode, scanning in Full MS mode (2 scans) from 60 to 900 m/z at 70,000 resolution, with 4 kV spray voltage, 45 sheath gas, 15 auxiliary gas. Calibration was performed prior to analysis using the PierceTM Positive and Negative Ion Calibration Solutions (ThermoFisher). Acquired data was then converted from raw to mzXML file format using Mass Matrix (Cleveland, OH). Samples were analyzed in randomized order with a technical mixture injected periodically through analysis to qualify instrument performance. Metabolite assignments, isotopologue distributions, and correction for expected natural abundances of deuterium, 13C, and 15N isotopes were performed using MAVEN (Princeton, NJ)
Ion Mode:NEGATIVE