MB Sample ID: SA194288

Local Sample ID:KL_3
Subject ID:SU002148
Subject Type:Mammal
Subject Species:Mus musculus
Taxonomy ID:10090
Genotype Strain:C57Bl/6 GEMMs
Age Or Age Range:Mutants: K: 90 days, KK: 60 days, KL and KKL: 40 days post Ad5-CMV-Cre
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003365 AN003366
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS
Column SeQuant ZIC-pHILIC (150 x 4.2mm,5um) SeQuant ZIC-pHILIC (150 x 4.2mm,5um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units

MS:

MS ID:MS003132
Analysis ID:AN003365
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:Mass spectrometer operated in full scan mode with positive and negative polarity switching at 35000 resolution at 200 m/z with detection range of 85 to 1, 275 m/z in full scan mode. Electro-spray ionization source (HESI) was set to 3.5 kV voltage for positive mode and 4.0 kV for negative mode, sheath gas was set to 50 and aux gas to 20 arbitrary units, capillary temperature 300 °C, probe heater temperature 120 °C.
Ion Mode:POSITIVE
Capillary Temperature:300 C
Capillary Voltage:4 kV
Collision Energy:NA
Collision Gas:NA
Ion Source Temperature:120 C
Mass Accuracy:1-2 ppm
Automatic Gain Control:1e6
Dataformat:profile
Acquisition Parameters File:Metabolomics_pHILIC_Parkville_v1.pdf
Analysis Protocol File:PQMS3-MBPF-WIN-0501_analysis.pdf
  
MS ID:MS003133
Analysis ID:AN003366
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:Mass spectrometer operated in full scan mode with positive and negative polarity switching at 35000 resolution at 200 m/z with detection range of 85 to 1, 275 m/z in full scan mode. Electro-spray ionization source (HESI) was set to 3.5 kV voltage for positive mode and 4.0 kV for negative mode, sheath gas was set to 50 and aux gas to 20 arbitrary units, capillary temperature 300 °C, probe heater temperature 120 °C.
Ion Mode:NEGATIVE
Capillary Temperature:300 C
Capillary Voltage:3.5 kV
Collision Energy:NA
Collision Gas:NA
Ion Source Temperature:120 C
Mass Accuracy:1-2 ppm
Automatic Gain Control:1e6
Dataformat:profile
Acquisition Parameters File:Metabolomics_pHILIC_Parkville_v1.pdf
Analysis Protocol File:PQMS3-MBPF-WIN-0501_analysis.pdf