Return to study ST002066 main page
MB Sample ID: SA194289
Local Sample ID: | KL_1 |
Subject ID: | SU002148 |
Subject Type: | Mammal |
Subject Species: | Mus musculus |
Taxonomy ID: | 10090 |
Genotype Strain: | C57Bl/6 GEMMs |
Age Or Age Range: | Mutants: K: 90 days, KK: 60 days, KL and KKL: 40 days post Ad5-CMV-Cre |
Gender: | Male and female |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003365 | AN003366 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | HILIC |
Chromatography system | Thermo Dionex Ultimate 3000 RS | Thermo Dionex Ultimate 3000 RS |
Column | SeQuant ZIC-pHILIC (150 x 4.2mm,5um) | SeQuant ZIC-pHILIC (150 x 4.2mm,5um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE | NEGATIVE |
Units |
MS:
MS ID: | MS003132 |
Analysis ID: | AN003365 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Mass spectrometer operated in full scan mode with positive and negative polarity switching at 35000 resolution at 200 m/z with detection range of 85 to 1, 275 m/z in full scan mode. Electro-spray ionization source (HESI) was set to 3.5 kV voltage for positive mode and 4.0 kV for negative mode, sheath gas was set to 50 and aux gas to 20 arbitrary units, capillary temperature 300 °C, probe heater temperature 120 °C. |
Ion Mode: | POSITIVE |
Capillary Temperature: | 300 C |
Capillary Voltage: | 4 kV |
Collision Energy: | NA |
Collision Gas: | NA |
Ion Source Temperature: | 120 C |
Mass Accuracy: | 1-2 ppm |
Automatic Gain Control: | 1e6 |
Dataformat: | profile |
Acquisition Parameters File: | Metabolomics_pHILIC_Parkville_v1.pdf |
Analysis Protocol File: | PQMS3-MBPF-WIN-0501_analysis.pdf |
MS ID: | MS003133 |
Analysis ID: | AN003366 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Mass spectrometer operated in full scan mode with positive and negative polarity switching at 35000 resolution at 200 m/z with detection range of 85 to 1, 275 m/z in full scan mode. Electro-spray ionization source (HESI) was set to 3.5 kV voltage for positive mode and 4.0 kV for negative mode, sheath gas was set to 50 and aux gas to 20 arbitrary units, capillary temperature 300 °C, probe heater temperature 120 °C. |
Ion Mode: | NEGATIVE |
Capillary Temperature: | 300 C |
Capillary Voltage: | 3.5 kV |
Collision Energy: | NA |
Collision Gas: | NA |
Ion Source Temperature: | 120 C |
Mass Accuracy: | 1-2 ppm |
Automatic Gain Control: | 1e6 |
Dataformat: | profile |
Acquisition Parameters File: | Metabolomics_pHILIC_Parkville_v1.pdf |
Analysis Protocol File: | PQMS3-MBPF-WIN-0501_analysis.pdf |