Return to study ST002263 main page
MB Sample ID: SA217396
Local Sample ID: | Klb_KO_IF_3 |
Subject ID: | SU002349 |
Subject Type: | Mammal |
Subject Species: | Mus musculus |
Taxonomy ID: | 10090 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003696 | AN003697 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | HILIC |
Chromatography system | Thermo Vanquish | Thermo Vanquish |
Column | Waters Acquity BEH Amide (150 x 2.1mm,1.7um) | Waters Acquity BEH Amide (150 x 2.1mm,1.7um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive HF hybrid Orbitrap | Thermo Q Exactive HF hybrid Orbitrap |
Ion Mode | POSITIVE | NEGATIVE |
Units | relative counts | relative counts |
MS:
MS ID: | MS003447 |
Analysis ID: | AN003696 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Full MS-ddMS2 data was collected, an inclusion list was used to prioritize MS2 selection of metabolites from our in-house ‘local’ library, when additional scan bandwidth was available MS2 was collected in a data-dependent manner. Mass range was 60-900 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.5 Da. Metabolomics data was processed using MS-DIAL v4.60. Features were excluded from analysis if peak height was not at least 5-fold greater in one or more samples compared to the procedural blank average. |
Ion Mode: | POSITIVE |
Analysis Protocol File: | DeFelice_Methods.docx |
MS ID: | MS003448 |
Analysis ID: | AN003697 |
Instrument Name: | Thermo Q Exactive HF hybrid Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Full MS-ddMS2 data was collected, an inclusion list was used to prioritize MS2 selection of metabolites from our in-house ‘local’ library, when additional scan bandwidth was available MS2 was collected in a data-dependent manner. Mass range was 60-900 mz, resolution was 60k (MS1) and 15k (MS2), centroid data was collected, loop count was 4, isolation window was 1.5 Da. Metabolomics data was processed using MS-DIAL v4.60. Features were excluded from analysis if peak height was not at least 5-fold greater in one or more samples compared to the procedural blank average. |
Ion Mode: | NEGATIVE |
Analysis Protocol File: | DeFelice_Methods.docx |