Analysis ID AN003926
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Vanquish
Column CSH C18 column (100 mm length × 2.1 mm i.d.; 1.7-µm particle size)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive HF-X Orbitrap
Ion Mode POSITIVE
Units peak height

MS:

MS ID:MS003664
Analysis ID:AN003926
Instrument Name:Thermo Q Exactive HF-X Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:HESI source conditions are as follows: sheath gas flow 55, auxiliary gas flow 15, sweep gas flow 3, capillary temperature 275°C, S-lens RF level 50, auxiliary gas heater temperature 450 °C, and needle voltage 3500 V and -3500 V for positive and negative ionization mode, respectively. DDA MS/MS spectra were acquired for the top 4 ions. MS scans were collected with 60k resolving power from 120-1700 m/z, AGC target of 106 ions, and maximum accumulation time of 100 ms. MS/MS spectra were collected with 15k resolving power, 1 Da isolation window, normalized collision energy of 20, 30, 60, 2 s dynamic exclusion window, 8×103 AGC target, and 50 ms maximum accumulation time. Spectra were stored in centroid mode. Three rounds of iterative exclusion MS/MS were acquired for each pooled QC sample.