Summary of Study ST000100

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000092. The data can be accessed directly via it's Project DOI: 10.21228/M81S38 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST000100
Study TitleGlobal Metabolomics of C. elegans using an augmented reference IROA design
Study Typeheat shock
Study SummaryGlobal Metabolomics of C. elegans using an augmented reference IROA design
Institute
University of Florida
DepartmentBiochemistry & Molecular Biology
LaboratoryArthur Edison
Last NameEdison
First NameArthur
AddressR3-226 Academic Research Building, Department of Biochemistry & Molecular Biology, PO Box 100245, Gainesville, FL 32610-0245
Emailaedison@ufl.edu
Submit Date2014-09-13
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Uploaded File Size16 G
Analysis Type DetailLC-MS
Release Date2015-02-03
Release Version1
Arthur Edison Arthur Edison
https://dx.doi.org/10.21228/M81S38
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN000165 AN000166
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS
Column Waters Acquity HSS T3 (100 x 2.1mm,1.8um) Waters Acquity HSS T3 (100 x 2.1mm,1.8um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units

Chromatography:

Chromatography ID:CH000114
Chromatography Summary:Samples were analyzed using a mass range of m/z 70-1000 in positive and negative ionization mode, externally calibrated, using a Thermo Scientific Q-Exactive Orbitrap mass spectrometer equipped with a Dionex UltiMate 3000 RS autosampler and pump. Three ?L of each sample were injected onto an Acquity UPLC HSS T3 (2.1x100 mm, 1.8 micron) column using a column temperature of 35 C and a flow rate of 300 L/min with solvent A (0.1% formic acid in water (Fisher LC-MS Optima)) held at 100% A for 1 min, followed by linear increase to 80% B (0.1% formic acid in acetonitrile) over 10 min, kept constant for 2 min, followed by a return to initial conditions over 0.5 min and equilibration for 3 min.
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Waters Acquity HSS T3 (100 x 2.1mm,1.8um)
Flow Gradient:A held at 100% A for 1 min, followed by linear increase to 80% B over 10 min, kept constant for 2 min, followed by a return to initial conditions over 0.5 min and equilibration for 3 min
Flow Rate:300 L/min
Sample Injection:3 ?L
Solvent A:100% water; 0.1% formic acid
Solvent B:100% acetonitrile; 0.1% formic acid
Chromatography Type:Reversed phase