Summary of Study ST001682
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001081. The data can be accessed directly via it's Project DOI: 10.21228/M8ZT4C This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST001682 |
Study Title | Untargeted urine LC-HRMS metabolomics profiling for bladder cancer binary outcome classification |
Study Summary | Two samples cohorts were analysed for bladder cancer biomarkers selection. Untargeted urine RP UPLC-HRMS metabolomics profiling was utilized in SCAN MS mode and positive polarity. Dilute and shoot technique was employed for sample preparation. |
Institute | Lomonosov MSU |
Department | Chemistry |
Laboratory | Mass spectrometry |
Last Name | Plyushchenko |
First Name | Ivan |
Address | Leninskie gory, 1/3 |
plyush1993@bk.ru | |
Phone | +79160362200 |
Submit Date | 2020-10-20 |
Num Groups | 2 |
Total Subjects | 124 |
Raw Data Available | Yes |
Raw Data File Type(s) | cdf |
Analysis Type Detail | LC-MS |
Release Date | 2021-02-22 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN002746 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Shimadzu 20AD |
Column | Waters Acquity BEH C18 (100 x 2mm,1.7um) |
MS Type | ESI |
MS instrument type | Other |
MS instrument name | Shimadzu LCMS-IT-TOF |
Ion Mode | POSITIVE |
Units | area |
Chromatography:
Chromatography ID: | CH002028 |
Chromatography Summary: | The apparatus consisted of a two LC-20AD pumps (Shimadzu, Japan), an automatic sample injector, coupled on-line with an LCMS-IT-TOF (Shimadzu, Japan) mass spectrometer with an electrospray ionization source. The HPLC separation was conducted on a 2.1 × 100 mm Waters Acquity UPLC BEH; 1.7 μm column (Waters, Milford MA, USA) with 2.1 × 5 mm UHPLC Guard 3PK Zorbax SB-C18 1.8 μm guard column (Agilent Technologies, USA). Two solvents were used: (A) deionized water with 0,1% formic acid, and (B) MeCN. After a 5 min hold at 5% solvent B, elution was performed according to the following conditions: from 5% B to 40% B in 20 min; to 95% B in 7 min; keeping constant for 10 min, to 5% B in 3 min, followed by 5 min of equilibration on the initial conditions. Total analysis time was 50 min. Column temperature was 35˚C. Autosampler tray temperature was 4˚C. |
Instrument Name: | Shimadzu 20AD |
Column Name: | Waters Acquity BEH C18 (100 x 2mm,1.7um) |
Column Pressure: | 300-500 bar |
Column Temperature: | 35 |
Flow Gradient: | non-linear |
Flow Rate: | 0.3 ml/min |
Injection Temperature: | 5 |
Internal Standard: | papaverine |
Sample Injection: | 5 mkl |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 100% acetonitrile |
Analytical Time: | 50 min |
Chromatography Type: | Reversed phase |