This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001418. The data can be accessed directly via it's Project DOI: 10.21228/M8F409 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002224
Study TitleIntracellular metabolic profile of renal cells cultured in Plasmax
Study SummaryThe objective of this experiment is to analyse the metabolic profiles of human renal epithelial cells HK2 and ccRCC cell lines 786-O, 786-M1A and 786-M2A in Plasmax media. The experiment was conducted on three different days using cells with different passage numbers. This is Part 5 of a study and the experimental number is MS55.
Institute
CECAD Research Center
Last NameYang
First NameMing
AddressJoseph-Stelzmann-Straße 26, Köln, Koeln, 50931, Germany
Emailming.yang@uni-koeln.de
Phone4922147884306
Submit Date2022-07-15
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2022-08-03
Release Version1
Ming Yang Ming Yang
https://dx.doi.org/10.21228/M8F409
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003633
Analysis type MS
Chromatography type HILIC
Chromatography system Thermo Dionex Ultimate 3000
Column SeQuant ZIC-pHILIC
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units peak area

Chromatography:

Chromatography ID:CH002688
Chromatography Summary:Chromatographic separation of metabolites was achieved using a Millipore Sequant ZIC-pHILIC analytical column (5 µm, 2.1 × 150 mm) equipped with a 2.1 × 20 mm guard column (both 5 mm particle size) with a binary solvent system. Solvent A was 20 mM ammonium carbonate, 0.05% ammonium hydroxide; Solvent B was acetonitrile. The column oven and autosampler tray were held at 40 °C and 4 °C, respectively. The chromatographic gradient was run at a flow rate of 0.200 mL/min as follows: 0–2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-23 min: hold at 80% B. Samples were randomized and the injection volume was 5 µl. A pooled quality control (QC) sample was generated from an equal mixture of all individual samples and analysed interspersed at regular intervals.
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:SeQuant ZIC-pHILIC
Column Temperature:40
Flow Gradient:0-2 min: 80% B; 2-17 min: linear gradient from 80% B to 20% B; 17-17.1 min: linear gradient from 20% B to 80% B; 17.1-23 min: hold at 80% B
Flow Rate:0.200 mL/min
Solvent A:100% water; 20 mM ammonium carbonate; 0.05% ammonium hydroxide
Solvent B:100% acetonitrile
Chromatography Type:HILIC