Summary of Study ST001371

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000937. The data can be accessed directly via it's Project DOI: 10.21228/M8K97X This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  |  Perform analysis on untargeted data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001371
Study TitleUntargeted metabolomics of Quercus ilex acorns
Study TypeLC-MSE analysis
Study SummaryUPLC-MSE analysis of samples from Quercus ilex acorns flour. The objective of the study is to obtain a metabolomic profile of several acorns from different trees. This phytochemical analysis and characterization will be a base for the identification of bioactive, antinutritional, or toxic compounds and traceability analysis.
Institute
Universidad de Córdoba
DepartmentDepartment Biochemistry and Molecular Biology
LaboratoryAgroforestry and Plant Biochemistry, Proteomics and Systems Biology
Last NameLópez-Hidalgo
First NameCristina
AddressCampus de Rabanales; Edificio C6, Planta Baja
Emailn12lohic@uco.es
Phone626894948
Submit Date2020-04-25
Raw Data AvailableYes
Raw Data File Type(s)mzXML
Analysis Type DetailLC-MS
Release Date2020-06-01
Release Version1
Cristina López-Hidalgo Cristina López-Hidalgo
https://dx.doi.org/10.21228/M8K97X
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002288 AN002289
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Waters Acquity H-Class Waters Acquity H-Class
Column Waters C18 (100 x 2.1mm,1.7um) Waters C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Waters Synapt G2 XS QTOF Waters Synapt G2 XS QTOF
Ion Mode POSITIVE NEGATIVE
Units abundance abundance

MS:

MS ID:MS002132
Analysis ID:AN002288
Instrument Name:Waters Synapt G2 XS QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:MS analysis was conducted in an ultra-performance liquid chromatography (UPLC Acquity H-Class, Waters, Milford, USA) coupled to a quadrupole time of flight (QTof) G2-XS mass spectrometer (Waters, Milford, USA). The chromatographic separation was carried out in a C18 column (2.1×100 mm, 1.7 µm, Waters, Milford, USA), kept at 45 °C. The injection volume was 5 µL, and the flow rate set at 0.450 mL min−1. Mobile phases consisted of 0.1 % formic acid in Milli-Q water (A) and methanol (B). The gradient elution profile was as follow (time, % B): 0 min, 2% B; 0.25 min, 2 % B; 12.25 min, 99 % B; 13.0 min, 99 % B; 13.01 min; 2 % B; 17.00 min; 2 % B and then the column was equilibrated for 5 min prior to each analysis. The MS acquisition was performed in negative and positive ionization modes in a scan range from m/z 100 to 1200 and time acquisition of 0 to 17 min. The analysis type performed was accurate mass screening on MSE data with a low collision energy of 4.00 eV and a high-energy ramp of 10.00 to 45.00 eV. The capillary and cone voltage were set at 2.50 kV and 40 V, respectively. The desolvation gas was set to 600 L h−1, the cone gas set to 50 L h−1 and the source and desolvation temperature was set to 100 °C and 250 °C, respectively. For automated accurate mass measurement, a solution of leucine-enkephalin (200 ng mL−1) in methanol: water (50:50) with 0.1% formic acid was used as lock mass and pumped at a flow rate of 5 µL min−1. The molecule of leucine-enkephalin (m/z 556.2766 in ESI+ and m/z 554.262 in ESI−) was used for recalibrating the mass axis and ensuring a robust accurate mass measurement at any time. For continuous quality assurance and to provide confidence in the data, quality control (QC), a mix prepared from equal volumes of all the samples was injected between every three samples in the batch along with methanol as a blank run to correct a drift of the raw signal intensity during the analysis. All the data acquired were exported by Waters UNIFI software in order to analyze by the software Progenesis QI (Nonlinear Dynamics, Newcastle, United Kingdom).
Ion Mode:POSITIVE
Collision Energy:4.00 eV
Fragment Voltage:High-ernergy ramp of 10.00 to 45.00 eV
Fragmentation Method:MSE
  
MS ID:MS002133
Analysis ID:AN002289
Instrument Name:Waters Synapt G2 XS QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:MS analysis was conducted in an ultra-performance liquid chromatography (UPLC Acquity H-Class, Waters, Milford, USA) coupled to a quadrupole time of flight (QTof) G2-XS mass spectrometer (Waters, Milford, USA). The chromatographic separation was carried out in a C18 column (2.1×100 mm, 1.7 µm, Waters, Milford, USA), kept at 45 °C. The injection volume was 5 µL, and the flow rate set at 0.450 mL min−1. Mobile phases consisted of 0.1 % formic acid in Milli-Q water (A) and methanol (B). The gradient elution profile was as follow (time, % B): 0 min, 2% B; 0.25 min, 2 % B; 12.25 min, 99 % B; 13.0 min, 99 % B; 13.01 min; 2 % B; 17.00 min; 2 % B and then the column was equilibrated for 5 min prior to each analysis. The MS acquisition was performed in negative and positive ionization modes in a scan range from m/z 100 to 1200 and time acquisition of 0 to 17 min. The analysis type performed was accurate mass screening on MSE data with a low collision energy of 4.00 eV and a high-energy ramp of 10.00 to 45.00 eV. The capillary and cone voltage were set at 2.50 kV and 40 V, respectively. The desolvation gas was set to 600 L h−1, the cone gas set to 50 L h−1 and the source and desolvation temperature was set to 100 °C and 250 °C, respectively. For automated accurate mass measurement, a solution of leucine-enkephalin (200 ng mL−1) in methanol: water (50:50) with 0.1% formic acid was used as lock mass and pumped at a flow rate of 5 µL min−1. The molecule of leucine-enkephalin (m/z 556.2766 in ESI+ and m/z 554.262 in ESI−) was used for recalibrating the mass axis and ensuring a robust accurate mass measurement at any time. For continuous quality assurance and to provide confidence in the data, quality control (QC), a mix prepared from equal volumes of all the samples was injected between every three samples in the batch along with methanol as a blank run to correct a drift of the raw signal intensity during the analysis. All the data acquired were exported by Waters UNIFI software in order to analyze by the software Progenesis QI (Nonlinear Dynamics, Newcastle, United Kingdom).
Ion Mode:NEGATIVE
Collision Energy:4.00 eV
Fragment Voltage:High-ernergy ramp of 10.00 to 45.00 eV
Fragmentation Method:MSE