Summary of Study ST001450

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000996. The data can be accessed directly via it's Project DOI: 10.21228/M8Z692 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  |  Perform analysis on untargeted data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001450
Study TitleFive Easy Metrics of Data Quality for LC-MS based Global Metabolomics
Study SummaryData quality in global metabolomics is of great importance for biomarker discovery and systems biology studies. However, comprehensive metrics and methods to evaluate and compare the data quality of global metabolomics data sets are lacking. In this work, we combine newly developed metrics, along with well-known measures, to comprehensively and quantitatively characterize the data quality across two similar LC-MS platforms, with the goal of providing an efficient and improved ability to evaluate the data quality in global metabolite profiling experiments. A pooled human serum sample was run 50 times on two high-resolution LC-QTOF-MS platforms to provide profile and centroid MS data. These data were processed using Progenesis Qi software and then analyzed using five important data quality measures, including retention time drift, compound coverage, missing values and MS reproducibility (2 measures). The coverage was fit to a Gamma distribution versus compound abundance, which was normalized to allow comparison of different platforms. To evaluate missing values, characteristic curves were obtained by plotting the compound detection percentage versus extraction frequency. To characterize reproducibility, the accumulative coefficient of variation (CV) versus percentage of total compounds detected and CV vs intra-class correlation coefficient (ICC) were investigated. Key findings include significantly better performance using profile mode data compared to centroid mode as well quantitatively better performance from the newer, higher resolution instrument. A summary of the results given in tabulated form gives a snapshot of the experimental results and provides a template to evaluate the global metabolite profiling workflow. In total, these measures give a good overall view of data quality in global profiling and allow comparisons of data acquisition strategies and platforms as well as optimization of parameters.
Institute
University of Washington
DepartmentAnesthesiology and Pain
LaboratoryDaniel Raftery
Last NameZhang
First NameXinyu
Address850 Republican Street, Seattle, Washington 98109, United States
Emailxinyu.z@live.com
Phone850-567-2757
Submit Date2020-08-18
Raw Data AvailableYes
Raw Data File Type(s)d
Analysis Type DetailLC-MS
Release Date2020-09-10
Release Version1
Xinyu Zhang Xinyu Zhang
https://dx.doi.org/10.21228/M8Z692
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002421 AN002422 AN002423 AN002424
Analysis type MS MS MS MS
Chromatography type HILIC HILIC HILIC HILIC
Chromatography system Agilent 1260 Infinity Agilent 1260 Infinity Agilent 1290 Infinity Agilent 1290 Infinity
Column Waters XBridge BEH Amide (15 cm x 2.1mm,2.5um) Waters XBridge BEH Amide (15 cm x 2.1mm,2.5um) Waters XBridge BEH Amide (15 cm x 2.1mm,2.5um) Waters XBridge BEH Amide (15 cm x 2.1mm,2.5um)
MS Type ESI ESI ESI ESI
MS instrument type QTOF QTOF QTOF QTOF
MS instrument name Agilent 6520 QTOF Agilent 6520 QTOF Agilent 6545 QTOF Agilent 6545 QTOF
Ion Mode POSITIVE POSITIVE POSITIVE POSITIVE
Units Peak area Peak area Peak area Peak area

MS:

MS ID:MS002258
Analysis ID:AN002421
Instrument Name:Agilent 6520 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:The ESI conditions were as follows: Electrospray ion source ESI Agilent Jet Stream Technology in positive ionization mode; voltage 3.8 kV; desolvation temperature 325 °C; cone flow 20 L/h; desolvation gas flow 600 L/h; nebulizer pressure 45 psi, N2 drying gas; MS scan rate of 1.03 spectra/s across the range m/z 60-1000. The software, Progenesis Qi, was used to process raw data. In the results, every compound (feature) had at least two ions, like an ion and its isotope ion or adduct ion. Compounds with single ions were filtered out. If a compound had m/z information only, the m/z was directly used in the compound name. If a compound had neutral mass information, the neutral mass was converted to m/z.
Ion Mode:POSITIVE
  
MS ID:MS002259
Analysis ID:AN002422
Instrument Name:Agilent 6520 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:The ESI conditions were as follows: Electrospray ion source ESI Agilent Jet Stream Technology in positive ionization mode; voltage 3.8 kV; desolvation temperature 325 °C; cone flow 20 L/h; desolvation gas flow 600 L/h; nebulizer pressure 45 psi, N2 drying gas; MS scan rate of 1.03 spectra/s across the range m/z 60-1000. The software, Progenesis Qi, was used to process raw data. In the results, every compound (feature) had at least two ions, like an ion and its isotope ion or adduct ion. Compounds with single ions were filtered out. If a compound had m/z information only, the m/z was directly used in the compound name. If a compound had neutral mass information, the neutral mass was converted to m/z.
Ion Mode:POSITIVE
  
MS ID:MS002260
Analysis ID:AN002423
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:(Agilent 6545) MS scan rate of 1.03 spectra/s across the range m/z 60-1000. The software, Progenesis Qi, was used to process raw data. In the results, every compound (feature) had at least two ions, like an ion and its isotope ion or adduct ion. Compounds with single ions were filtered out. If a compound had m/z information only, the m/z was directly used in the compound name. If a compound had neutral mass information, the neutral mass was converted to m/z.
Ion Mode:POSITIVE
  
MS ID:MS002261
Analysis ID:AN002424
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:(Agilent 6545) MS scan rate of 1.03 spectra/s across the range m/z 60-1000. The software, Progenesis Qi, was used to process raw data. In the results, every compound (feature) had at least two ions, like an ion and its isotope ion or adduct ion. Compounds with single ions were filtered out. If a compound had m/z information only, the m/z was directly used in the compound name. If a compound had neutral mass information, the neutral mass was converted to m/z.
Ion Mode:POSITIVE