Summary of Study ST002162

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001375. The data can be accessed directly via it's Project DOI: 10.21228/M8069C This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002162
Study TitleCFAP418 participates in membrane-associated cellular processes through binding lipids during ciliogenesis
Study SummaryCiliopathies and retinal degenerative diseases are heterogeneous groups of genetic diseases. CFAP418 is a causative gene of both diseases, and its sequence is evolutionarily conserved. Here, we employ affinity purification coupled with mass spectrometry and quantitative lipidomic, proteomic, and phosphoproteomic approaches to address the function of CFAP418 in retinas. We show CFAP418 unexpectedly binds to lipid metabolism precursor phosphatidic acid (PA) and mitochondrion-specific lipid cardiolipin but does not form a tight and static complex with proteins. Loss of Cfap418 leads to a widespread disruption of membrane lipid homeostasis and changes in protein-membrane association, which subsequently causes mitochondrial morphological and functional defects and membrane remodeling abnormalities in multiple vesicular trafficking pathways. The signaling of PA-binding protein kinase Ca is increased. Our results indicate that membrane lipid imbalance is a new pathological mechanism underlying inherited ciliopathies and retinal degenerations, which is associated with other known causative RAB28 and BBS genes.
Institute
University of Utah - Metabolomics Core
Last NameMaschek
First NameJohn
AddressEmma Eccles Jones Medical Science Building, 15 N Medical Dr East, Salt Lake City, UT, 84112, USA
Emailalan.maschek@pharm.utah.edu
Phone801-587-7779
Submit Date2022-05-10
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2023-05-10
Release Version1
John Maschek John Maschek
https://dx.doi.org/10.21228/M8069C
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN003542 AN003543
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Agilent 6550 Agilent 6550
Column Waters Acquity CSH C18 (100 x 2.1mm,1.7um) Waters Acquity CSH C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type QTOF QTOF
MS instrument name Agilent 6540 QTOF Agilent 6545 QTOF
Ion Mode POSITIVE NEGATIVE
Units pmol per sample pmol per sample

MS:

MS ID:MS003300
Analysis ID:AN003542
Instrument Name:Agilent 6540 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:For positive mode, the source gas temperature was set to 225 °C, with a drying gas flow of 11 L/minute, nebulizer pressure of 40 psig, sheath gas temp of 350 °C and sheath gas flow of 11 L/minute. VCap voltage is set at 3500 V, nozzle voltage 500V, fragmentor at 110 V, skimmer at 85 V and octopole RF peak at 750 V. For data processing, Agilent MassHunter (MH) Workstation and software packages MH Qualitiative and MH Quantitative were used. The pooled QC (n=8) and process blank (n=4) were injected throughout the sample queue to ensure the reliability of acquired lipidomics data. For lipid annotation, accurate mass and MS/MS matching was used with the Agilent Lipid Annotator library. Results from the positive and negative ionization modes from Lipid Annotator were merged based on the class of lipid identified. Data exported from MH Quantitative was evaluated using Excel where initial lipid targets are parsed based on the following criteria. Only lipids with relative standard deviations (RSD) less than 30% in QC samples are used for data analysis. Additionally, only lipids with background AUC counts in process blanks that are less than 30% of QC are used for data analysis. The parsed excel data tables are normalized based on the ratio to class-specific internal standards, then to sum prior to statistical analysis.
Ion Mode:POSITIVE
  
MS ID:MS003301
Analysis ID:AN003543
Instrument Name:Agilent 6545 QTOF
Instrument Type:QTOF
MS Type:ESI
MS Comments:For negative mode, the source gas temperature was set to 300 °C, with a drying gas flow of 11 L/minute, a nebulizer pressure of 30 psig, sheath gas temp of 350 °C and sheath gas flow 11 L/minute. VCap voltage was set at 3500 V, nozzle voltage 75 V, fragmentor at 175 V, skimmer at 75 V and octopole RF peak at 750 V. For data processing, Agilent MassHunter (MH) Workstation and software packages MH Qualitiative and MH Quantitative were used. The pooled QC (n=8) and process blank (n=4) were injected throughout the sample queue to ensure the reliability of acquired lipidomics data. For lipid annotation, accurate mass and MS/MS matching was used with the Agilent Lipid Annotator library. Results from the positive and negative ionization modes from Lipid Annotator were merged based on the class of lipid identified. Data exported from MH Quantitative was evaluated using Excel where initial lipid targets are parsed based on the following criteria. Only lipids with relative standard deviations (RSD) less than 30% in QC samples are used for data analysis. Additionally, only lipids with background AUC counts in process blanks that are less than 30% of QC are used for data analysis. The parsed excel data tables are normalized based on the ratio to class-specific internal standards, then to sum prior to statistical analysis.
Ion Mode:NEGATIVE