Summary of Study ST002681

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001666. The data can be accessed directly via it's Project DOI: 10.21228/M8C13S This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002681
Study TitleNon-T2D vs T2D
Study SummaryPlasma samples from Senegalese individuals with T2D (n=31) or without T2D (n=34) were compared using mass-spectrometry-based metabolomics analyses.
Institute
University of Colorado Anschutz Medical Campus
Last NameNemkov
First NameTravis
Address12801 E 17th Avenue, RC-1 South, Rm 9403G, Aurora, CO, 80045, USA
Emailtravis.nemkov@cuanschutz.edu
Phone303-724-3253
Submit Date2023-04-25
Raw Data AvailableYes
Raw Data File Type(s)mzXML
Analysis Type DetailLC-MS
Release Date2023-10-26
Release Version1
Travis Nemkov Travis Nemkov
https://dx.doi.org/10.21228/M8C13S
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004353 AN004354
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Phenomenex Kinetex C18 (100 x 2.1mm,1.7um) Phenomenex Kinetex C18 (100 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units AU AU

MS:

MS ID:MS004100
Analysis ID:AN004353
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The Q Exactive mass spectrometer (ThermoFisher) was operated independently in positive or negative ion mode, scanning in Full MS mode (2 microscans) from 60 to 900 m/z at 70,000 resolution, with 4 kV spray voltage, 45 sheath gas, 15 auxiliary gas. Calibration was performed prior to analysis using the PierceTM Positive and Negative Ion Calibration Solutions (ThermoFisher). Acquired data was then converted from raw to mzXML file format using Mass Matrix (Cleveland, OH). Samples were analyzed in randomized order with a technical mixture injected periodically through analysis to qualify instrument performance. Metabolite assignments, isotopologue distributions, and correction for expected natural abundances of deuterium, 13C, and 15N isotopes were performed using MAVEN (Princeton, NJ)
Ion Mode:POSITIVE
  
MS ID:MS004101
Analysis ID:AN004354
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The Q Exactive mass spectrometer (ThermoFisher) was operated independently in positive or negative ion mode, scanning in Full MS mode (2 microscans) from 60 to 900 m/z at 70,000 resolution, with 4 kV spray voltage, 45 sheath gas, 15 auxiliary gas. Calibration was performed prior to analysis using the PierceTM Positive and Negative Ion Calibration Solutions (ThermoFisher). Acquired data was then converted from raw to mzXML file format using Mass Matrix (Cleveland, OH). Samples were analyzed in randomized order with a technical mixture injected periodically through analysis to qualify instrument performance. Metabolite assignments, isotopologue distributions, and correction for expected natural abundances of deuterium, 13C, and 15N isotopes were performed using MAVEN (Princeton, NJ)
Ion Mode:NEGATIVE
  logo