This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001078. The data can be accessed directly via it's Project DOI: 10.21228/M8C111 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  |  Perform analysis on untargeted data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001676
Study TitleLipidomic analysis of CD4+ T-cell subsets (Th1,Th2,Th17 and iTreg cells) (part I)
Study TypeMS, untargeted cell-based lipidomics
Study SummaryPart 1/5: It includes lipidomic analysis of CD4+ T-cell subsets(Th1,Th2,Th17 and iTreg cells)and their paired controls(Th0 cells).
Institute
University of Turku
DepartmentSystems Medicine, Turku Bioscience
LaboratorySystems Medicine
Last NameSen
First NamePartho
AddressTykistökatu 6B, BioCity, 5th Floor, Turku, Southwest, 20521, Finland
Emailpartho.sen@utu.fi
Phone0469608145
Submit Date2021-01-28
Raw Data AvailableYes
Raw Data File Type(s)mzML
Analysis Type DetailLC-MS
Release Date2021-11-02
Release Version1
Partho Sen Partho Sen
https://dx.doi.org/10.21228/M8C111
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002734
Analysis type MS
Chromatography type Reversed phase
Chromatography system Waters Acquity UHPLC UNSPSC 41115709
Column Waters BEH C18 (00mm x 2.1mm,1.7um )
MS Type ESI
MS instrument type QTOF
MS instrument name Agilent 6540 QTOF
Ion Mode POSITIVE
Units Intensities

Chromatography:

Chromatography ID:CH002020
Chromatography Summary:Chromatographic separation was performed on an ACQUITY UHPLC BEH C18 column (2.1 mm × 100 mm, particle size 1.7 µm, Waters, Milford, MA, USA). The flow rate was set at 0.4 ml/min throughout the run with an injection volume of 1 µL. The following solvents were used for the gradient elution: Solvent A was H2O with 1% NH4Ac (1M) and HCOOH (0.1%) added. Solvent B was a mixture of ACN: IPA (1:1 v/v) with 1% NH4Ac (1M) and HCOOH (0.1%) added. The gradient was programmed as follows: 0 to 2 min 35-80% B, 2 to 7 min 80-100 % B, 7 to 14 min 100% B. The column was equilibrated with a 7min period of 35 % B prior to the next run.
Methods Filename:parthosen_20210128_003828_PR_CH_Chromatography_metadata.pdf
Instrument Name:Waters Acquity UHPLC UNSPSC 41115709
Column Name:Waters BEH C18 (00mm x 2.1mm,1.7um )
Flow Gradient:The gradient was programmed as follows: 0 to 2 min 35-80% B, 2 to 7 min 80-100 % B, 7 to 14 min 100% B. The column was equilibrated with a 7min period of 35 % B prior to the next run.
Flow Rate:0.4ml/min
Solvent A:100% water; 0.1% formic acid; 1% ammonium acetate
Solvent B:50% acetonitrile/50% isopropanol; 0.1% formic acid; 1% ammonium acetate
Chromatography Type:Reversed phase