Return to study ST003024 main page
MB Sample ID: SA327770
Local Sample ID: | M_Cefto3g_191_2 |
Subject ID: | SU003138 |
Subject Type: | Bacteria |
Subject Species: | Pseudomonas aeruginosa |
Taxonomy ID: | 287 |
Genotype Strain: | CW41 |
Age Or Age Range: | NA |
Weight Or Weight Range: | NA |
Height Or Height Range: | NA |
Gender: | Not applicable |
Cell Strain Details: | hypermutable clinical isolate |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004958 | AN004959 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | HILIC |
Chromatography system | Thermo Dionex Ultimate 3000 | Thermo Dionex Ultimate 3000 |
Column | Merck SeQuant ZIC-pHILIC (150 x 4.6mm,5um) | Merck SeQuant ZIC-pHILIC (150 x 4.6mm,5um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE | NEGATIVE |
Units | peak height | peak height |
MS:
MS ID: | MS004698 |
Analysis ID: | AN004958 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Full scan LCMS data acquired using polarity switching. raw data processed using IDEOM workflow ( 10.1007/978-1-0716-0239-3_21). |
Ion Mode: | POSITIVE |
Capillary Temperature: | 300 C |
Capillary Voltage: | 4 kV |
Collision Energy: | NA |
Collision Gas: | NA |
Dry Gas Flow: | 50 |
Dry Gas Temp: | 120 |
Ion Source Temperature: | 120 C |
Ionization: | ESI |
Mass Accuracy: | 3 ppm |
Precursor Type: | [M+H]+ |
Source Temperature: | 300 |
Acquisition Parameters File: | Metabolomics_pHILIC_Parkville_v1.pdf |
Analysis Protocol File: | PQMS3-MPMF-WIN-0501_LCMS_data_acquisition_for_untargeted_metabolomics_analysis.pdf |
MS ID: | MS004699 |
Analysis ID: | AN004959 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Full scan LCMS data acquired using polarity switching. raw data processed using IDEOM workflow ( 10.1007/978-1-0716-0239-3_21). |
Ion Mode: | NEGATIVE |
Capillary Temperature: | 300 C |
Capillary Voltage: | 3.5 kV |
Collision Energy: | NA |
Collision Gas: | NA |
Dry Gas Flow: | 50 |
Dry Gas Temp: | 120 |
Ion Source Temperature: | 120 C |
Ionization: | ESI |
Mass Accuracy: | 3 ppm |
Precursor Type: | [M-H]- |
Source Temperature: | 300 |
Acquisition Parameters File: | Metabolomics_pHILIC_Parkville_v1.pdf |