Summary of Study ST002104
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001333. The data can be accessed directly via it's Project DOI: 10.21228/M8DM7G This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST002104 |
Study Title | Chemoresistant Cancer Cell Lines are Characterized by Migratory, Amino Acid Metabolism, Protein Catabolism and IFN1 Signalling Perturbations |
Study Type | Biomedical research |
Study Summary | Our analysis was able to separate chemoresistant cells from their parental cells based on their metabolomic and proteomic features and identified altered biological processes and pathways which are of further interest. Preliminary investigation of patient-derived cells highlighted the need to perform broad biological and molecular analyses, compre-hensive in vitro and in vivo studies, using a larger patient cohort to achieve a deeper and clinically relevant characterization of the molecular drivers of chemoresistance. |
Institute | Future Industries Institute |
Laboratory | Manuela Klingler-Hoffmann |
Last Name | Acland |
First Name | Mitchell |
Address | X Building, Mawson Lakes South Australia 5095 |
mitch.acland@gmail.com | |
Phone | 0448671141 |
Submit Date | 2022-02-06 |
Num Groups | 4 |
Total Subjects | 12 |
Num Males | NA |
Num Females | NA |
Study Comments | OVCAR5 and CaOV3 cell lines and their carboplatin resistant cell lines |
Publications | Chemoresistant Cancer Cell Lines are Characterized by Migra-tory, Amino Acid Metabolism, Protein Catabolism and IFN1 Signalling Perturbations |
Raw Data Available | Yes |
Raw Data File Type(s) | raw(Thermo) |
Analysis Type Detail | LC-MS |
Release Date | 2022-03-29 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003439 | AN003440 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | HILIC | HILIC |
Chromatography system | Thermo Dionex Ultimate 3000 RS | Thermo Dionex Ultimate 3000 RS |
Column | SeQuant ZIC-pHILIC (150 x 4.2mm,5um) | SeQuant ZIC-pHILIC (150 x 4.2mm,5um) |
MS Type | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE | NEGATIVE |
Units | peak height | peak height |
Chromatography:
Chromatography ID: | CH002541 |
Chromatography Summary: | LCMS data was acquired on Q-Exactive Orbitrap mass spectrometer (Thermo Fisher) coupled with high-performance liquid chromatography (HPLC) system Dionex Ultimate® 3000 RS (Thermo Fisher). Chromatographic separation was performed on a ZIC-pHILIC column (5 µm, polymeric, 150 × 4.6 mm, SeQuant®, Merck). The mobile phase (A) was 20 mM ammonium carbonate and (B) acetonitrile. The gradient program started at 80% B and was reduced to 50% B over 15 min, then reduced from 50% B to 5% B over 3 min, followed by wash with 5% B for another 3 min, and finally 8 min re-equilibration with 80% B. The flow rate was 0.3 mL/min and column compartment temperature was 40ºC. The total run time was 32 min with an injection sample volume of 10 µL. |
Methods Filename: | Metabolomics_pHILIC_Parkville_v1.meth |
Instrument Name: | Thermo Dionex Ultimate 3000 RS |
Column Name: | SeQuant ZIC-pHILIC (150 x 4.2mm,5um) |
Column Pressure: | 60 bar at starting conditions |
Column Temperature: | 25 C |
Flow Gradient: | 0 min - 80%B, 15 min - 50%B, 18 min - 5%B, 21 min - 5%B, 24 min - 80%B, 32 min - 80%B |
Flow Rate: | 0.3 ml/min |
Injection Temperature: | 4 C |
Internal Standard: | CAPS, CHAPS, PIPES |
Sample Injection: | 10 uL |
Solvent A: | 100% water; 20 mM ammonium formate |
Solvent B: | 100% acetonitrile |
Analytical Time: | 32 min |
Capillary Voltage: | 3.5 kV |
Oven Temperature: | 25 C |
Washing Buffer: | syringe wash 50% IPA |
Weak Wash Solvent Name: | 10% MeOH |
Strong Wash Solvent Name: | 10% MeOH |
Sample Loop Size: | 25 uL |
Sample Syringe Size: | 25 uL |
Chromatography Type: | HILIC |