Return to study ST000083 main page
MB Sample ID: SA004260
Local Sample ID: | SBEP_Microbiome.004 |
Subject ID: | SU000102 |
Subject Type: | Animal |
Subject Species: | Mus musculus |
Taxonomy ID: | 10090 |
Genotype Strain: | 129/SvJ |
Age Or Age Range: | 6- to 8-week-old |
Gender: | Female |
Animal Animal Supplier: | Jackson Laboratories,Bar Harbor, ME |
Animal Housing: | specific pathogen-free conditions in filter-top cages |
Animal Feed: | food provided ad libitum |
Animal Water: | sterile water |
Species Group: | Mammals |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN000135 |
---|---|
Analysis type | MS |
Chromatography type | GC |
Chromatography system | Agilent 7890A |
Column | Agilent HP5-MS (30m × 0.25mm, 0.25 um) |
MS Type | EI |
MS instrument type | Single quadrupole |
MS instrument name | Agilent 5975C |
Ion Mode | POSITIVE |
Units | Peak area |
Chromatography:
Chromatography ID: | CH000095 |
Chromatography Summary: | Agilent 7890A gas chromatograph with a HP-5MS gas chromatography column using Chemstation |
Chromatography Comments: | Chromatography was carried out on an Agilent 7890A gas chromatograph using the manufacturer's software (Chemstation) and a HP-5MS gas chromatography column (Agilent Technologies, Santa Clara, CA; 30 m x 0.25 mm x 0.25 m film thickness). The sample injection mode was splitless, and 1 L of each sample was injected. The injection port temperature was held at 250 C throughout the analysis. The GC oven was held at 60 C for 1 min after injection, and the temperature was then increased to 325 C by 10 C/min, followed by a 5 min hold at 325 C. The helium gas flow rates for each Experiment were determined by the Agilent Retention Time Locking function based on analysis of deuterated myristic acid and were in the range of 0.450.5 mL/min. |
Instrument Name: | Agilent 7890A |
Column Name: | Agilent HP5-MS (30m × 0.25mm, 0.25 um) |
Flow Rate: | 0.450.5 mL/min |
Injection Temperature: | 250 C |
Sample Injection: | 1 L, splitless |
Analytical Time: | 37.5 min |
Oven Temperature: | 60 C for 1 min, then increased to 325 C by 10 C/min, followed by a 5 min hold at 325 C |
Sample Syringe Size: | 10 L |
Chromatography Type: | GC |