Summary of Study ST002506

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001619. The data can be accessed directly via it's Project DOI: 10.21228/M8F41P This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST002506
Study TitleNatural abundance of isotopic metabolite detection in mouse eye orgnaoids.
Study SummaryBecause the natural abundance of isotopic metabolites in the early eye organoids has not yet been reported, we have performed LC-MS/MS without exogenous isotopic labeling to show the natural abundance of isotopic carbons. Cell Name AES0145 : Rx-GFP K/I EB5 (RIKEN Cell Bank): Organoid method (PMID: 21475194).
Institute
Northwestern University
Last NameTAKATA
First NameNOZOMU
Address303 East Superior Street, 10-220
Emailnozomu.takata@northwestern.edu
Phone3125036066
Submit Date2023-03-15
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2023-05-01
Release Version1
NOZOMU TAKATA NOZOMU TAKATA
https://dx.doi.org/10.21228/M8F41P
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN004128
Analysis type MS
Chromatography type HILIC
Chromatography system Q-exactive
Column Waters XBridge BEH Amide (100 x 3.0mm, 3.5um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode UNSPECIFIED
Units Peak area

MS:

MS ID:MS003875
Analysis ID:AN004128
Instrument Name:Thermo Q Exactive Orbitrap
Instrument Type:Orbitrap
MS Type:ESI
MS Comments:The capillary of the ESI source was set to 275 °C, with sheath gas at 35 arbitrary units, auxiliary gas at 5 arbitrary units and the spray voltage at 4.0 kV. In positive/negative polarity switching mode, an m/z scan range from 60 to 900 was chosen and MS1 data was collected at a resolution of 70,000. The automatic gain control (AGC) target was set at 1 × 106 and the maximum injection time was 200 ms. The top 5 precursor ions were subsequently fragmented, in a data-dependent manner, using the higher energy collisional dissociation (HCD) cell set to 30% normalized collision energy in MS2 at a resolution power of 17,500. Besides matching m/z, metabolites are identified by matching either retention time with analytical standards and/or MS2 fragmentation pattern. Data acquisition and analysis were carried out by Xcalibur 4.1 software and Tracefinder 4.1 software, respectively (both from Thermo Fisher Scientific)
Ion Mode:UNSPECIFIED