This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001722. The data can be accessed directly via it's Project DOI: 10.21228/M8441W This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST002764 |
Study Title | Identification of pre-diagnostic lipid sets associated with liver cancer risk using untargeted lipidomics and chemical set analysis – a nested case-control study within the ATBC cohort |
Study Summary | In pre-disposed individuals, a reprogramming of the hepatic lipid metabolism may support liver cancer initiation. We conducted a high-resolution mass spectrometry based untargeted lipidomics analysis of pre-diagnostic serum samples from a nested case-control study (219 liver cancer cases and 219 controls) within the Alpha-Tocopherol, Beta-Carotene Cancer Prevention (ATBC) Study. Out of 462 annotated lipids, 158 (34.2%) were associated with liver cancer risk in a conditional logistic regression analysis at a false discovery rate (FDR) < 0.05. A chemical set enrichment analysis (ChemRICH) and co-regulatory set analysis suggested that 22/28 lipid classes and 47/83 correlation modules were significantly associated with liver cancer risk (FDR <0.05). Strong positive associations were observed for monounsaturated fatty acids (MUFA), triacylglycerols (TAGs), and phosphatidylcholines (PCs) having MUFA acyl chains. Negative associations were observed for sphingolipids (ceramides and sphingomyelins), lysophosphatidylcholines, cholesterol esters and polyunsaturated fatty acids (PUFA) containing TAGs and PCs. Stearoyl-CoA desaturase enzyme 1 (SCD1), a rate limiting enzyme in fatty acid metabolism and ceramidases seems to be critical in this reprogramming. In conclusion, our study reports pre-diagnostic lipid changes that provide novel insights into hepatic lipid metabolism reprogramming may contribute to a pro-cell growth and anti-apoptotic tissue environment and, in turn, support liver cancer initiation. Study |
Institute | Icahn School of Medicine at Mount Sinai |
Department | Environmental Medicine and Public Health |
Laboratory | Integrated Data Science Laboratory for Metabolomics and Exposomics |
Last Name | Barupal |
First Name | Dinesh |
Address | CAM Building 102street |
dinesh.barupal@mssm.edu | |
Phone | 5309794354 |
Submit Date | 2023-06-28 |
Raw Data Available | Yes |
Raw Data File Type(s) | mzML |
Analysis Type Detail | LC-MS |
Release Date | 2023-07-03 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004498 | AN004499 |
---|---|---|
Analysis type | MS | MS |
Chromatography type | Reversed phase | Reversed phase |
Chromatography system | Agilent 1290 Infinity | Agilent 1290 Infinity |
Column | Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um) | Waters ACQUITY UPLC BEH C18 (100 x 2.1mm,1.7um) |
MS Type | ESI | ESI |
MS instrument type | QTOF | QTOF |
MS instrument name | Agilent 6550 QTOF | Agilent 6530 QTOF |
Ion Mode | NEGATIVE | POSITIVE |
Units | normalized peak intensity | normalized peak intensity |
MS:
MS ID: | MS004245 |
Analysis ID: | AN004498 |
Instrument Name: | Agilent 6550 QTOF |
Instrument Type: | QTOF |
MS Type: | ESI |
MS Comments: | The quadrupole/time-of-flight (QTOF) mass spectrometers are operated with electrospray ionization (ESI) performing full scan in the mass range m/z 65–1700 in positive (Agilent 6530, equipped with a JetStreamSource) and negative (Agilent 6550, equipped with a dual JetStream Source) modes producing both unique and complementary spectra. Instrument parameters are as follows (positive mode) Gas Temp 325°C, Gas Flow 8 l/min, Nebulizer 35 psig, Sheath Gas 350°C, Sheath Gas Flow 11, Capillary Voltage 3500 V, Nozzle Voltage 1000V, Fragmentor 120V, Skimmer 65V. Data (both profile and centroid) are collected at a rate of 2 scans per second. In negative ion mode, Gas Temp 200°C, Gas Flow 14 l/min, Fragmentor 175V, with the other parameters identical to positive ion mode. For the 6530 QTOF, a reference solution generating ions of 121.050 and 922.007 m/z in positive mode and 119.036 and 966.0007 m/z in negative mode, and these are used for continuous mass correction. For the 6550, the reference solution is introduced via a dual spray ESI, with the same ions and continuous mass correction. Samples are injected (1.7 μl in positive mode and 5 μl in negative ion mode) with a needle wash for 20 seconds (wash solvent is isopropanol). The valve is switched back and forth during the run for washing; this has been shown to be essential for reducing carryover of less polar lipids. |
Ion Mode: | NEGATIVE |
MS ID: | MS004246 |
Analysis ID: | AN004499 |
Instrument Name: | Agilent 6530 QTOF |
Instrument Type: | QTOF |
MS Type: | ESI |
MS Comments: | The quadrupole/time-of-flight (QTOF) mass spectrometers are operated with electrospray ionization (ESI) performing full scan in the mass range m/z 65–1700 in positive (Agilent 6530, equipped with a JetStreamSource) and negative (Agilent 6550, equipped with a dual JetStream Source) modes producing both unique and complementary spectra. Instrument parameters are as follows (positive mode) Gas Temp 325°C, Gas Flow 8 l/min, Nebulizer 35 psig, Sheath Gas 350°C, Sheath Gas Flow 11, Capillary Voltage 3500 V, Nozzle Voltage 1000V, Fragmentor 120V, Skimmer 65V. Data (both profile and centroid) are collected at a rate of 2 scans per second. In negative ion mode, Gas Temp 200°C, Gas Flow 14 l/min, Fragmentor 175V, with the other parameters identical to positive ion mode. For the 6530 QTOF, a reference solution generating ions of 121.050 and 922.007 m/z in positive mode and 119.036 and 966.0007 m/z in negative mode, and these are used for continuous mass correction. For the 6550, the reference solution is introduced via a dual spray ESI, with the same ions and continuous mass correction. Samples are injected (1.7 μl in positive mode and 5 μl in negative ion mode) with a needle wash for 20 seconds (wash solvent is isopropanol). The valve is switched back and forth during the run for washing; this has been shown to be essential for reducing carryover of less polar lipids. |
Ion Mode: | POSITIVE |