Summary of Study ST001027

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000685. The data can be accessed directly via it's Project DOI: 10.21228/M84H56 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  |  Perform analysis on untargeted data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001027
Study TitleInfluence of Data-Processing Strategies on Normalized Lipid Levels using an Open-Source LC-HRMS/MS Lipidomics Workflow
Study SummaryLipidomics is an emerging field with significant potential for improving clinical diagnosis and our understanding of health and disease. While the diverse biological roles of lipids contribute to their clinical utility, the unavailability of lipid internal standards representing each species, make lipid quantitation analytically challenging. The common approach is to employ one or more internal standards for each lipid class examined and use a single point calibration for normalization (relative quantitation). To aid in standardizing and automating this relative quantitation process, we developed LipidMatch Normalizer (LMN) http://secim.ufl.edu/secim-tools/ which can be used in most open source lipidomics workflows. While the effect of lipid structure on relative quantitation has been investigated, applying LMN we show that data-processing can significantly affect lipid semi-quantitative amounts. Polarity and adduct choice had the greatest effect on normalized levels; when calculated using positive versus negative ion mode data, one fourth of lipids had greater than 50 % difference in normalized levels. Based on our study, sodium adducts should not be used for statistics when sodium is not added intentionally to the system, as lipid levels calculated using sodium adducts did not correlate with lipid levels calculated using any other adduct. Relative quantification using smoothing versus not smoothing, and peak area versus peak height, showed minimal differences, except when using peak area for overlapping isomers which were difficult to deconvolute. By characterizing sources or variation introduced during data-processing and introducing automated tools, this work helps increase through-put and improve data-quality for determining relative changes across groups.
Institute
University of Florida
DepartmentChemistry
LaboratoryRichard Yost Laboratory
Last NameLevy
First NameAllison
Address214 Leigh Hall, PO Box 117200, Gainesville, Florida, 32611, USA
Emailallisonjlevy@ufl.edu
Phone3523920515
Submit Date2018-07-25
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2018-08-27
Release Version1
Allison Levy Allison Levy
https://dx.doi.org/10.21228/M84H56
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN001684 AN001685
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Dionex Ultimate 3000 RS Thermo Dionex Ultimate 3000 RS
Column Waters Acquity BEH C18 (150 x 2.1mm,1.7um) Waters Acquity BEH C18 (150 x 2.1mm,1.7um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units peak area peak area

Chromatography:

Chromatography ID:CH001185
Chromatography Summary:Liquid Chromatography Protocol Samples were injected onto a Waters (Milford, MA) BEH C18 UHPLC column (50 x 2.1 mm, 1.7 µm) held at 50 °C with mobile phase A consisting of acetonitrile:water (60:40, v/v) with 10 mM ammonium formate and 0.1% formic acid and mobile phase B consisting of isopropanol:acetonitrile:water (90:8:2) with 10 mM ammonium formate and 0.1% formic acid at a flow rate of 0.5 mL/min. A Dionex Ultimate 3000 RS UHLPC system (Thermo Scientific, San Jose, CA) coupled to a Thermo Q-Exactive mass spectrometer (San Jose, CA) was employed for data acquisition. The UHPLC gradient use in this experiment is shown in Table 1. Time (min) C (%) D (%) 0 80 20 1 80 20 3 70 30 4 55 45 6 40 60 8 35 65 10 35 65 15 10 90 17 2 98 18 2 98 19 80 20 23 80 20 Table 1: Gradient for reverse phase liquid chromatography of lipids. Mobile phase C consisted of 60:40 acetonitrile:water and mobile phase D consisted of 90:8:2 isopropanol:acetonitrile:water, with both containing 0.1% formic acid 10 mM ammonium formate. The flow rate was 500 µL/min.
Instrument Name:Thermo Dionex Ultimate 3000 RS
Column Name:Waters Acquity BEH C18 (150 x 2.1mm,1.7um)
Column Temperature:50
Flow Gradient:Time (min) C (%) D (%) 0 80 20 1 80 20 3 70 30 4 55 45 6 40 60 8 35 65 10 35 65 15 10 90 17 2 98 18 2 98 19 80 20 23 80 20
Flow Rate:0.5 mL/min
Solvent A:60% acetonitrile/40% water; 0.1% formic acid; 10 mM ammonium formate
Solvent B:90% isopropanol/8% acetonitrile/2% water; 0.1% formic acid; 10 mM ammonium formate
Chromatography Type:Reversed phase