Summary of Study ST001177

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000789. The data can be accessed directly via it's Project DOI: 10.21228/M8PT1R This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001177
Study TitleAnalysis of short chain phosphatidylcholine (PC) on the Golgi membrane
Study TypeGolgi membrane lipids characterization
Study SummaryStudies on vesicle formation by the Coat Protein I (COPI) complex have contributed to a basic understanding of how vesicular transport is initiated. We have identified that short chain lipids promote membrane properties that are conducive for fission. Here we investigated short chain PCs on Golgi membrane. These findings will advance the understanding of how lipid geometry contributes to membrane deformation needed for vesicle fission.
Institute
University of Kentucky
Last NameMorris
First NameAndrew
Address900 S limestone
Emaila.j.morris@uky.edu
Phone859-323-3749
Submit Date2019-04-25
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2019-05-15
Release Version1
Andrew Morris Andrew Morris
https://dx.doi.org/10.21228/M8PT1R
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN001953
Analysis type MS
Chromatography type Reversed phase
Chromatography system Thermo Dionex Ultimate 3000
Column ACQUITY BEH C8 column (100 x 2.1mm,1.7um)
MS Type ESI
MS instrument type Orbitrap
MS instrument name Thermo Q Exactive Orbitrap
Ion Mode POSITIVE
Units Peak area

Chromatography:

Chromatography ID:CH001416
Chromatography Summary:Lipid extracts were separated on a Waters ACQUITY BEH C8 column (2.1 × 100 mm, 1.7 μm) with the temperature maintained at 40 °C. The flow rate was 250 μL/min, and the mobile phases were consisted of 60:40 water/acetonitrile (A), and 90:10 isopropanol/acetonitrile (B), both containing 10 mM ammonium formate and 0.1% formic acid. The samples were eluted with a linear gradient from 5 % B to 97 % B over 20 min, maintained at 97 % B for 4 min and re-equilibration with 5 % B for 6 min. The sample injection volume was 5 μL
Instrument Name:Thermo Dionex Ultimate 3000
Column Name:ACQUITY BEH C8 column (100 x 2.1mm,1.7um)
Column Temperature:40
Flow Gradient:a linear gradient from 5 % B to 97 % B over 20 min, maintained at 97 % B for 4 min and re-equilibration with 5 % B for 6 min.
Flow Rate:250 µL/min
Solvent A:60% water/40% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Solvent B:90% isopropanol/10% acetonitrile; 0.1% formic acid; 10 mM ammonium formate
Chromatography Type:Reversed phase