Summary of Study ST000891

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000620. The data can be accessed directly via it's Project DOI: 10.21228/M8J098 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Show all samples  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST000891
Study TitleNMR comparison of urine samples by 1D NOESY presat and PURGE
Study TypeNMR development for metabolomics
Study SummaryDifferent water suppression pulse sequences, namely 1D NOESY presat and optimized PURGE were compared for urine samples. The original PURGE sequence was also tested to show the postential for lineshape distortions with the pulse sequence.
Institute
University of Georgia
DepartmentComplex Carbohydrate Research Center
LaboratoryComplex Carbohydrates Research Center
Last NameEdison
First NameArthur
Address315 Riverbend Road 30602 Athens GA
Emailaedison@uga.edu
Phone+1-706-542-8156
Submit Date2017-09-01
Num Groups1
Total Subjects10
Raw Data AvailableYes
Raw Data File Type(s)fid
Analysis Type DetailNMR
Release Date2017-11-15
Release Version1
Arthur Edison Arthur Edison
https://dx.doi.org/10.21228/M8J098
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


NMR:

NMR ID:NM000109
Analysis ID:AN001453
Instrument Name:Bruker Avance III
Instrument Type:FT-NMR
NMR Experiment Type:1D 1H
NMR Comments:4 spectra per sample: the first is the noesypr1d, then an optimized version of PURGE, then noesygppr1d, and finally the original PURGE (called zgpurge by Bruker)
Field Frequency Lock:2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz
Spectrometer Frequency:600 MHz
NMR Probe:TCI 600 H-C/N-D | TCI 600 H-C/N-D | TCI 600 H-C/N-D | TCI 600 H-C/N-D
NMR Solvent:H2O+D2O (90/10) potassium buffer | H2O+D2O (90/10) potassium buffer | H2O+D2O (90/10) potassium buffer | H2O+D2O (90/10) potassium buffer
NMR Tube Size:5 mm | 5 mm | 5 mm | 5 mm
Shimming Method:Topshim | Topshim | Topshim | Topshim
Pulse Sequence:noesypr1d | zgpurge-gprefocus-4 | noesygppr1d | zgpurge
Water Suppression:Presaturation | Presaturation | Presaturation | Presaturation
Pulse Width:19.94 μL | 19.94 μL | 19.94 μL | 19.94 μL
Power Level:5.5584 W | 5.5584 W | 5.5584 W | 5.5584 W
Receiver Gain:90.5 | 90.5 | 90.5 | 90.5
Offset Frequency:2824.58 Hz | 2824.58 Hz | 2824.58 Hz | 2824.58 Hz
Presaturation Power Level:0.000088411 W | 0.000088411 W | 0.000088411 W | 0.000088411 W
Chemical Shift Ref Cpd:TSP | TSP | TSP | TSP
Temperature:300 K | 300 K | 300 K | 300 K
Number Of Scans:32 | 32 | 32 | 32
Dummy Scans:4 | 4 | 4 | 4
Acquisition Time:2.726 | 2.726 | 2.726 | 2.726
Relaxation Delay:4 s | 4 s | 4 s | 4 s
Spectral Width:20.03 ppm | 20.03 ppm | 20.03 ppm | 20.03 ppm
Num Data Points Acquired:65536 | 65536 | 65536 | 65536
Real Data Points:32768 | 32768 | 32768 | 32768
Line Broadening:0.3 Hz | 0.3 Hz | 0.3 Hz | 0.3 Hz
Zero Filling:131072 | 131072 | 131072 | 131072
Apodization:Gaussian | Gaussian | Gaussian | Gaussian
Baseline Correction Method:Polynomial order 2 | Polynomial order 2 | Polynomial order 2 | Polynomial order 2
Chemical Shift Ref Std:DSS | DSS | DSS | DSS