Summary of Study ST001644

This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001052. The data can be accessed directly via it's Project DOI: 10.21228/M8QD76 This work is supported by NIH grant, U2C- DK119886.

See: https://www.metabolomicsworkbench.org/about/howtocite.php

Perform statistical analysis  |  Show all samples  |  Show named metabolites  |  Download named metabolite data  
Download mwTab file (text)   |  Download mwTab file(JSON)   |  Download data files (Contains raw data)
Study IDST001644
Study TitleIn Vitro Characterization and Metabolomic Analysis of Cold-Stored Platelets
Study SummaryPlatelet concentrates are currently stored at room temperature (RPs) under constant agitation for up to 5-7 days depending on national regulations. However, platelet quality deteriorates during storage and room temperature storage also increases the risk of bacterial growth. Previous studies have shown that cold-stored platelets (CPs) have higher hemostatic function and can be stored for up to three weeks. While these studies have compared the metabolic phenotypes of CPs and RPs, they have not compared the impact of storage temperature and cold agitation (CPAs) on platelet function, nor have they identified metabolic correlates to such parameters. In vitro analysis showed CPAs and CPs had reduced count, faster CD62P expression and increased lactadherin binding. Furthermore, CPAs and CPs had higher maximal aggregation and a reduced aggregation lag phase compared to RPs. Metabolomic analysis revealed CPAs and CPs exhibited lower oxidative stress shown by preserved glutathione and pentose phosphate pools. CPAs and CPs also had reduced markers of beta-oxidation and amino acid catabolism demonstrating reduced needs for energy. Agitation did not significantly impact in vitro function or metabolomic parameters of cold-stored platelets. Correlation of in vitro and metabolomic results highlighted important metabolites that may contribute to stored platelet functions.
Institute
University of Colorado Anschutz Medical Campus
DepartmentBiochemistry and Molecular Genetics
LaboratoryAngelo D'Alessandro
Last NameD'Alessandro
First NameAngelo
Address12801 E 17th Ave L18-9403D
Emailangelo.dalessandro@cuanschutz.edu
Phone3037245798
Submit Date2021-01-08
Num Groups3
Total Subjects8
Raw Data AvailableYes
Raw Data File Type(s)raw(Thermo)
Analysis Type DetailLC-MS
Release Date2021-01-25
Release Version1
Angelo D'Alessandro Angelo D'Alessandro
https://dx.doi.org/10.21228/M8QD76
ftp://www.metabolomicsworkbench.org/Studies/ application/zip

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002689 AN002690
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system Thermo Vanquish Thermo Vanquish
Column Phenomenex Kinetex C18 (150 x 2.1mm,2.6um) Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode NEGATIVE POSITIVE
Units Relative Abundance Relative Abundance

Chromatography:

Chromatography ID:CH001983
Chromatography Summary:UHPLC-MS metabolomics analyses were performed as described,(1, 3) using a Vanquish UHPLC system coupled online to a high-resolution Q Exactive mass spectrometer (Thermo Fisher, Bremen, Germany). Samples were resolved over a Kinetex C18 column (2.1x150 mm, 1.7 µm; Phenomenex, Torrance, CA, USA) at 45°C. A volume of 10 μl of sample extracts was injected into the UHPLC-MS. Each sample was injected and run with two different chromatographic and MS conditions using a 5 minute gradient at 450 µL/minute from 5-95% B (A: 5% acetonitrile, 95%water/1 mM ammonium acetate; B:95%acetonitrile/5% water, 1 mM ammonium acetate) and the MS was operated in negative ion mode.
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Column Temperature:45
Flow Gradient:5 minute gradient from 5-95% B
Flow Rate:450 µL/min
Solvent A:5% acetonitrile/95% water; 1 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 1 mM ammonium acetate
Chromatography Type:Reversed phase
  
Chromatography ID:CH001984
Chromatography Summary:UHPLC-MS metabolomics analyses were performed as described,(1, 3) using a Vanquish UHPLC system coupled online to a high-resolution Q Exactive mass spectrometer (Thermo Fisher, Bremen, Germany). Samples were resolved over a Kinetex C18 column (2.1x150 mm, 1.7 µm; Phenomenex, Torrance, CA, USA) at 45°C. A volume of 10 μl of sample extracts was injected into the UHPLC-MS. Each sample was injected and run with two different chromatographic and MS conditions, using a 5 minute gradient at 450 µL/minute from 5-95% B (A: water/0.1% formic acid; B:acetonitrile/0.1% formic acid) and the MS was operated in positive mode.
Instrument Name:Thermo Vanquish
Column Name:Phenomenex Kinetex C18 (150 x 2.1mm,2.6um)
Column Temperature:45
Flow Gradient:5 minute gradient from 5-95% B
Flow Rate:450 µL/min
Solvent A:5% acetonitrile/95% water; 1 mM ammonium acetate
Solvent B:95% acetonitrile/5% water; 1 mM ammonium acetate
Chromatography Type:Reversed phase